"Time Remain" is stored even if the pump was switched OFF.
NOTE:
"Time Remain" Will automatically adjust if Total Volume is changed
during operation.
If TIME function is OFF, in technician mode, the TIME function will
not be available.
"Time Remain." Will not display if total time is more than 100 hours.
"VTBI" Display
1. Pressing the SELECT key twice during operation causes
the displayed rate to change for 5 seconds to left "VTBI"
"VOLUME INFUSED" Display.
1. Pressing the SELECT key during operation causes the
displayed rate to change for 5 seconds to
the accumulated volume infused.
The VTBI and INFUSED volume will automatically reset to zero,
NOTE:
once it reaches 9999 ml.
The volume INFUSED, can be zeroed only using the CLEAR key.
When a program was completed the VTBI will reset to the original
setting
Changing Infusion Rate During Operation
1. Press the RUN/HOLD key to halt the operation of the
pump, HOLD will appear on display.
NOTE: Use the Select key to display current data : Rate,
Time left, VTBI, Infused volume.
3. Set the desired infusion rate using the arrow keys.
Pressing continuously will increase the speed and steps of
changes.
4. Press the RUN/HOLD key to renew operation.
NIKI V4 Operator's Manual
21
VTBI
RATE
VI
RATE-
-
VOL.-