Changing the current default "Pressure" Limit to another
Press SELECT key continuously.
Switch on the pump by using the ON/OFF key.
Press the RUN/HOLD key, PRES shows on display for one
second, thereafter the current pressure will appear.
By pressing the SELECT key scroll through, and choose the
desired limit.
NOTE
The following pressure parameters can be set:
Low
Normal = 0.5 bar.
High
NOTE
The selected pressure setting will remain in the pump's memory until
changed again.
Press the RUN/HOLD key, in order to store the parameter in the pump's memory.
NOTE
The pump has an antibolus system, which prevent a bolus delivery
to the patient, once a downstream occlusion is released.