Pressure Detection Test
Each pump is designed to detect an occlusion downstream of the pump.
If a pump fails to detect a downstream occlusion, the device should be removed from
service and repaired. Perform the following steps at room temperature to ensure proper
operation of the downstream occlusion sensor.
1. Load and prime an administration set.
2. Start the pump at a rate of 125 ml/hr and run for 1 minute.
3. Fold and pinch the tubing just below the pump.
The "HI PRSS" message for HIGH PRESSURE should appear on the display, the
audible alarm should sound and the pump should stop.
WARNING:
Upstream Occlusion Test
The device is designed to detect an occluded administration set.
Perform the following steps to ensure the performance of the upstream occlusion
sensor.
1. Load and prime an administration set.
2. Pinch off the tubing with a set clamp above the pump head.
3. Set rate of 125 ml/hr and start the pump.
4. Within 3 minutes the following should occur:
a. The pump stops.
b. The message AIR appears on the display.
c. The audible alarm is activated.
WARNING:
Charge Indicator Test
1.
Disconnect the power cord from the AC power outlet.
2.
Verify that the CHARGE indicator is off. After one second
the display starts flashing constantly to indicate that the pump
is operating on battery power.
3.
Connect the power cord to the AC power outlet. Verify that the
CHARGE indicator is on, and the display has stopped flashing.
NIKI V4 Operator's Manual
If the pump does not respond as described, remove the device
from service. Troubleshoot in accordance with section 4.4 and
repair in accordance with section4.5.
If these indications do not occur, remove the device from service.
Troubleshoot in accordance with section 4.4 and repair in accordance
with section 4.5.