c. Pressure setting
(code 3)
Set code 3. The display will show the actual pressure. Should be between 80 to 120.
Press run/hold key to continue. The display will show high pressure level. This
number should be between 13 to 22.
Press run/hold key to continue and set the high pressure absolute level. This level
should be set to 250.
d. Clock setting
Set code 5. The display will show clock on/off. Alternate the actual setting with the
select key. The clock off choice will eliminate the use of the time setting.
d. Volume calibration setting
(code 1)
Set code 1. The display will show the actual calibration between 150 to 165. Those
numbers are the motor pulses requested to deliver 1 ml.
1.
Press continuously on the down arrow key while switching the pump on.
1.1 Press code 1 and press run key.
1.2 Pump will display "cal 1" massage for Baxter set and "cal 3" massage for Abbot
set.
1.3 If calibration is not suitable for the set you are working with change setting:
1.3.1 Turn the pump off and press continuously on the down arrow key while switching
the pump on again.
1.3.2 Set code 6 and press run key. Change set setting and turn the pump off.
2.
Turn the pump on. Set the pump to infuse at rate 200 ml/hr. VTBI 30 ml. Infuse
into a measuring cup.
3.
Scale the VI.
4.
If different than 30 (±5%) change calibration as follows:
4.1 Press continuously on the down arrow key while switching the pump on.
4.2
Set code 1.
4.3
Pump will display "cal" message press run key.
4.4
Pump will display a number – change according to the differences in volume (if
volume was more than VTBI, set to lower number and if the volume was lower
set to higher number).
4.5
Repeat steps 2-4 until VI will be 30 ml. ±5%.
NIKI V4 Operator's Manual