Tailoring the Pump to the Hospital Needs
Changing the current default Program to another:
1. Press SELECT key continuously.
2. Switch on the pump using the ON/OFF key.
3. Release the SELECT key once SLCT shows on display
Using the SELECT key, scroll through the different
programs and choose the desired program from the
following choices:
MICRO
5. Press the RUN/HOLD key, in order to store the parameter in
the pump's memory.
The selected program will remain in the pump's memory until changed
NOTE
again.
List of programs
The NIKI V4 volumetric infusion pump features 2 different
programs.
Following are the specifications for each program:
MACRO program.
Rate: 1 to 999ml/hr
Total volume to be infused: 1 to 9999ml.
MICRO program to be used in Neonatal and Pediatric wards.
Rate: 0.1 to 99.9 ml/hr in 0.1ml increments.
Total volume to be infused: 0.1 to 2500 ml.
NIKI V4 Operator's Manual
MACRO
34
RATE-
-
VOL.-
MACRO
MICRO