High Pressure (Occlusion) Alarm Test:
1. Occlude the administration set. Press the RUN/HOLD key to
start operation at a rate of 200 ml/hr.
.
2. After 10 seconds HI PRESSURE alarm will activate.
(at a rate of 60 ml/hr the high pressure alarm will occur after
60 sec. at a rate of 1 ml/hr after 13 min.)
3. Press RUN/HOLD key to silence the alarm.
4. HOLD appears on the display.
The pressure threshold can be changed to suit the ward need, (see
NOTE
Changing the current default "Pressure" Limit to another on page 32).
NIKI V4 Operator's Manual