2.1

Description

The Simple Logger
recording device with an alkaline battery pack. Line tracking is performed
such that 64 samples over one line cycle are taken. Frequency tracking is
performed over the range of ±2Hz around the nominal line frequency (50
or 60Hz). Harmonic measurements are calculated from these 64 samples
(Harmonics are only available from the Simple Logger
within the DataView
The Simple Logger
second. The measurement process is performed no more than eight times
per second. TRMS calculations are performed on a single line cycle. This
means that the input(s) are ignored between measurement intervals.
The main advantage of the logger is its ability to perform a variety of
recording tasks with easy and intuitive setup from a computer using
DataView
software.
®
Analog information on the input is sampled and converted to a digital
signal. This digital signal is processed and stored along with scale and
time information. An optically isolated Universal Serial Bus (USB) port
provides for the transfer of data from the instrument's internal memory to
the computer for analysis.
Simple Logger
II Model CL601
®
Shop for AEMC products online at:
CHAPTER 2

PRODUCT FEATURES

II Model CL601 is a battery powered, one channel
®
application).
®
II records TRMS at a rate of up to eight times per
®
ShopAEMC