4.10 Reset Switch Operation

The RESET switch is located in the battery compartment. To access it,
remove the battery compartment cover, then reset the instrument by
inserting a small tool (such as a paper clip) through the hole in the case
to activate it.
NOTE: It is recommended to only press the RESET switch when the logger stops
responding to a normal press button control when not connected to DataView
is not recommended to reset the instrument when the logger is recording, down-
loading or being configured.
If the logger is not responding to a button press, make sure the unit has
sufficient battery power. If the battery voltage is below 1.7V, the unit will not
respond to a button press. In this condition, pressing the RESET switch
will not restart the unit.
It is recommended to download any desired session, then erase the
memory before starting a new recording.
NOTE: The resumption of the logger operation in the above situations assumes
that the RESET switch cleared the fault(s). The logger will not resume normal
operation if the fault condition still exists.
The instrument will try to recover normally. However, under certain condi-
tions, the clock and memory full state may be reset.