Event: Recording with Memory Cleared
When a recording starts, the logger will continue to record until one of the
following occurs:
•
The Session is complete.
•
The Memory is full and the recording mode is Start/Stop.
•
The
PRESS button is pushed until the STOP (ORANGE) LED
lights up and is released before the next LED lights.
•
The Stop Recording command from the SLII Control Panel is sent
to the unit.
•
The battery voltage falls below 1.7V.
Event: Recording with a Partial or Full Memory
If the RED LED is double-blinking, the memory must be cleared before
any further recording can be performed.
If the RED LED is single-blinking prior to starting a New Recording Ses-
sion, the memory is partially full.
To save, clear or check memory availability, use the Simple Logger
Control Panel software.
There may be instances where the GREEN LED is also double-blinking
indicating that the logger is still recording. The user can choose to stop the
Recording Session and download the saved session(s) and/or erase the
memory.
NOTE: The logger memory cannot be erased while in the Record mode. The
recording must be stopped first.
Event: Memory Filled During Recording Session
If the logger is recording using the Start/Stop mode and memory is filled
before the Recording Session has finished, the session will end.
The following happens:
•
The GREEN LED single-blinks.
•
The RED LED double-blinks.
At this time:
•
The memory can be downloaded and erased.
•
A new recording can be started or scheduled once the memory is
erased.
16