10. Flow Handpiece: Unscrew and remove the cartridge nut from the
Figure 6
Flow
Handpiece
11. Heat Shield: Place the heat shield over the cannula and handpiece.
Figure 7
Heat Shield
& Bending Tool
12. Bending Tool: Use the bending tool to place a smooth radius curve
13. The device is now ready for operation.
10
handpiece. See Fig. 6.
Cartridge
Nut
Turn the unit 'OFF' before inserting or replacing a cartridge.
Insert a cartridge into the handpiece, cannula facing outward.
Slip cartridge nut over cannula and screw cap on clockwise lightly
- do not over-tighten.
NOTE: If the cartridge isn't properly seated into the handpiece,
turn the unit on and press the
control panel. The plunger must be in its retracted position to
accept the cartridge. When fully seated, a few millimeters of the
cartridge will be visible.
NOTE: Take care when replacing a cartridge during a procedure,
the handpiece, cartridge nut and the expended cartridge will be
hot to the touch.
Then turn the heat shield until the Flow indicator is visible. See Fig. 7.
Heat
Shield
on the cannula so it can extend to within 5 mm of the working length
of the canal. See Fig. 7.
Place the cannula between the two raised bending posts. Gently
bend the cannula to the desired angle.
Warning: Using fingers or any other device other than the Bending
Tool to curve the cartridge cannula can result in cartridge rupture or
handpiece damage.
Warning: Do not touch hot EHPs or hot cartridges.
button on the right side of the
Bending
Tool