3. Handpieces: Clean the exterior of the handpieces with a soft cloth
moistened with mild non-chlorinated detergent or disinfecting
solution.
Do not use disinfecting solutions which contain phenol, anti-
corrosives, very acidic or very alkaline solutions - do not
submerge handpieces.
4. The handpiece cables are color-coded and keyed to only fit the
proper receptacle on the console.
With the arrow on the connector facing up, gently push the
connector into the console receptacle. See Fig. 4
5. Place handpieces into their respective holders - Pack on the left side
and Flow on the right side.
6. Power Cord: Attach the power cord to the back of the console and
plug into a grounded electrical outlet.
7. Heat Shield: The heat shield (included) is designed to reduce the
Flow handpiece surface temperature. If the heat shield is not utilized,
avoid contact with the forward portion of the handpiece.
Sterilize the heat shield before first use and between each patient,
using steam autoclave for 15 minutes at 132°C.
8. Electric Heat Pluggers (EHP): Sterilize using steam autoclave for
15 minutes at 132°C. See Fig. 5.
9. Pack Handpiece: Push EHP into handpiece tip and slowly rotate it
until it clicks in with handpiece tip. EHP must be fully engaged to
achieve desired heat settings. See Fig. 5.
EHP
Gently
Rotate
And Push
EHP Into
Handpiece
Arrow
On
Connector
EHP
Fully
Engaged
Figure 4
Handpiece
Cable
Connection
Figure 5
Attaching
EHP
9