Cone Fit & Plugger Selection - DENTSPLY Clamus DUAL Manual

3d obturation system
Table of Contents

Advertisement

Cone-Fit & Plugger Selection
1. Prepare the canal optimally recognizing shaping facilitates 3D
cleaning and filling.
2. Select a larger size Calamus
and effectively over a range of a few millimeters in the coronal one-
third of the canal.
3. Select a medium size Calamus
passively and effectively over a range of a few millimeters in the
middle one-third of the canal.
4. Select a smaller size Calamus
passively, effectively and deeper in the straightaway portion of the
canal and to within 4-5 mm of the canal terminus.
5. Select the Calamus
through the straightaway portion of the canal and optimally to within
5 mm from the working length. Set the silicone stop at this depth to
promote safety and accuracy.
6. Fit a nonstandardized, fully tapered master cone in a fluid-filled canal
that visually goes to the working length, exhibits apical tugback and
is confirmed radiographically.
7. Dry the canal with appropriately sized paper points to determine final
working length.
8. Trim the master cone back to the canal terminus based on the paper
point drying technique.
9. Lubricate the master cone lightly with sealer and gently insert it to
length.
®
manual plugger that will work passively
®
manual plugger that will work
®
manual plugger that will work
®
Electric Heat Plugger (EHP) that will passively fit
23