Handpieces can be removed from the console for repair or
replacement. Gently pull the connection shield back (1) while
simultaneously pulling the handpiece connector from the
®
Calamus
Dual console (2). See Fig. 12
(1)
Gently Pull
Connection
Shield
Back

Cartridge Nut

Allow the FLOW handpiece to cool before removing the cartridge nut
and any excess gutta-percha.
Steam autoclave for 15 minutes at 132°C.

Heat Shield

Steam autoclave for 15 minutes at 132 °C.

Cartridges

Cartridges are for single patient use.
Prior to using a cartridge on a patient, wipe the cannula with alcohol
or a disinfectant.
Make sure that the device has been switched off and cooled down.
Store cartridges at room temperature.
Do not immerse the cartridges into any liquid.
Dispose of cartridge in a sharps receptacle.
Do not use cartridges after the expiration date.
Cartridges CANNOT be re-heated and/or re-used.
WARNING: Re-heating or re-using a cartridge can result in the
cartridge rupturing, which will permanently damage the FLOW
handpiece! Re-heating cartridges is a misuse of the Calamus
and voids the device warranty.
Figure 12
Removing
Handpieces
(2)
Pull Handpiece
Connector
from Console
®
Dual
19