Figure 43 — Display Test Results
Once all six display conditions have been verified confirm completion of a successful test by selecting "Test Passed"
button which will create a "Display Test passed" status.
Fan Test
Figure 44 — Fan Test
To execute the Fan Test select the "Start Test" button. The fan will cycle through four operation speeds while monitoring
the fan RPM for the correct reference value.
Figure 45 — Fan Test Results
Once the fan test has completed the RDX 8000 will report a pass/fail status and the user may click the "OK" button to log
the result.
Cartridge Access Test
The Cartridge Access Test checks the ability of each slot to properly insert/load, identify/communicate with RDX media,
and unload/eject RDX media. The unit will be taken offline from the iSCSI connection to prevent concurrent data access
attempts from the host system. Also at the start of the test all currently inserted RDX media will be bulk ejected from the
unit. It is suggested the test operator use a known good cartridge which can be used to test each slot.
66
RDX 8000 User Guide