Bayer HealthCare Bai An Jie 2 User Manual page 25

Hide thumbs Also See for Bai An Jie 2:
Table of Contents

Advertisement

What You See
What It Means
Low Battery symbol
fl ashes if battery power
is low.
Battery life has ended.
Temperature is outside
the operating range of
the meter (too hot or
too cold).
Meter does not
recognize the test disc.
Test strip was removed
during the test, or the
release button was
pressed during the test.
The meter was open
when the meter handle
was pulled.
(with continuous
beep)
44
Bayer Customer Service: (852) 8100 6386
What You Should Do
Replace the battery (CR2032)
within 20 readings or one week.
Replace the battery (CR2032).
Allow the meter to reach room
temperature.
Remove the test disc and inspect
the meter for foreign material.
Rotate the disc and reinstall it. If
error repeats, try inserting a new
disc. If error continues, contact
Bayer Customer Service.
Take care to not press the release
button or remove the test strip
during the test. Release the used
test strip and repeat the test (see
page 10). If the error continues,
contact Bayer Customer Service.
Close the meter and push the
handle in.
What You See
What It Means
1. You may have
applied blood too
early to the test strip.
2. There may be
moisture in the test
strip slot.
There was a problem
completely fi lling the
test strip; accurate test
result is not possible.
1. Meter has detected
an error; accurate
result is not possible.
2. The 10-test disc may
be past its expiration
date or may have
been damaged due
to intense heat or
moisture exposure.
Your blood glucose level
is very high or the meter
has detected an error;
accurate test result is
not possible.
www.bayerdiabetes.com
What You Should Do
1. Repeat the test and wait until
you see the "apply blood"
prompts before you touch the
strip to the blood drop.
2. Wick the moisture away (see
page 36), then repeat the test.
Release the used test strip and
repeat the test with a new strip
(see page 10).
1. Release the used test strip and
repeat the test with a new strip
(see page 10).
2. Replace the 10-test disc, using a
new box if necessary.
Turn the meter off, then on using
. Retest. If the error appears
again, call your physician or
healthcare professional. Your
blood glucose may be above
83.3 mmol/L. If the problem is
with the meter, contact Bayer
Customer Service.
45

Advertisement

Table of Contents
loading