Service and maintenance

10.1

Calibration

The device has been calibrated by the manufacturer before shipping. Recalibration is not
required provided the environmental conditions are maintained.
10.2

Cleaning

Clean the display and the keypad periodically. Use a dry cloth for this.
CAUTION
Damage due to detergents
Detergents can damage the device. Do not use detergents.
10.3

Firmware updates

The SENTRON PAC3100 supports firmware updates.
Use the SENTRON powerconfig software, Version V2.1 or higher, to install all updates. For
update instructions, please see the related documentation.
The update function is password protected.
The last devices settings made remain unchanged.
Repeating the firmware update after cancellation
NOTICE
Cancellation of a firmware update puts the device out of commission
Please ensure there is uninterrupted power supply. Ensure the configuration software
completes the update properly.
The device loses the current firmware if the update operation is cancelled. The device
cannot function without firmware. It is necessary to repeat the update to restore functionality
of the device.
After cancellation of an update operation, the firmware of the device can no longer be read
out. The configuration software must therefore obtain the currently set communication
parameters of the device from another source, e.g. by manual input.