Trigger and List Control Keys
The Trigger Control key controls output transient triggers. The List key controls the generation of output
lists. A list can contain up to 100 points, each of which can specify an output change (or transient). Refer
to Chapter 4 of the Programming Guide for more information about programming triggers and lists.
Trigger
Press this key to access the trigger control menu list.
Control
Display
INIT:IMMED
INIT:CONT <char>
TRIG:SOUR <char>
DELAY <value>
ABORT
SYNC:SOUR <char>
SYNC:PHASE <value>
Shift
List
Press this key to access the list commands
Display
COUNT <value>
DWEL:<index> <value>
FREQ:<index> <value>
FSLW:<index> <value>
IPK:<index> <value>
OFFS:<index> <value>
OSLW:<index> <value>
PHASE:<index> <value>
SHAP:<index> <char>
STEP<char>
TTLT:<index> <value>
VOLT:<index> <value>
VSLW:<index> <value>
Notes:
1
User-defined waveshapes will also appear in this list when created.
value = a numeric value
char = a character string parameter
index = a numeric value that represents a list point from 0 to 99
p
and
¯
and
p
Index
end of the list is reached. When a value is edited, pressing
advances to the next list point.
presently displayed list point.
Command Function
Initiate the transient trigger sequence immediately.
Set continuous trigger initiation (ON or OFF).
Select transient trigger source (BUS, EXT, TTLT or
IMM).
Set trigger delay in seconds.
Abort all trigger sequences.
Select trigger sync source (PHASE or IMM).
Set synchronous phase reference angle in degrees.
Command Function
Specifies the number of times a list repeats.
List of output dwell times.
List of output frequencies.
List of output frequency slew rates
List of output peak current limits.
List of dc output voltages.
Dc offset voltage slew rate list
List of output voltage phase angles.
List of output waveform shapes.
(SINE, SQUARE or CSIN) CSIN = clipped sine wave
Response of list to triggers (ONCE or AUTO).
List of Trigger Out pulses (0=no pulse; 1=pulse).
List of ac output voltages.
List of output voltage slew rates.
q
scroll through the command list.
°
scroll through the parameter list.
q
and
Index
scroll through the desired list points. EOL is displayed when the
P
ressing
Front Panel Operation - 4
1
automatically
Enter
truncates or clears the list at the
Clear Entry