Agilent Technologies 6811B User Manual page 60

Ac power solutions
Hide thumbs Also See for 6811B:
Table of Contents

Advertisement

A - Specifications
Table A-1. Performance Specifications (continued)
Parameter
Measurement Accuracy
(continued)
Harmonic Measurement
Accuracy
(50/60 Hz, @25 o C ±5 o C),
± (% of output + offset)
1 Specifications subject to change without notice.
2 These specifications are subject to the restrictions of Table 1-3.
3 Product may be operated between dc and 45 Hz subject to operating conditions described in Table A-3.
60
rms Current High Range
(45−100 Hz):
(>100−500 Hz):
(>500 Hz−1 kHz):
rms Current Low Range
(45−100 Hz):
(>100−500 Hz):
(>500 Hz−1 kHz):
repetitive pk current High Range
(45 Hz−1 kHz):
repetitive pk current Low Range
(45 Hz−1 kHz):
Power (VA) Low Range
(45−100 Hz):
(>100−500 Hz):
(>500 Hz-1 kHz):
Power (VA) High Range
(45−100 Hz):
(>100−500 Hz):
(>500 Hz−1 kHz):
Power (Watts) Low Range
(45−100 Hz):
(>100−500 Hz):
(>500 Hz−1 kHz):
Power (Watts) High Range
(45−100 Hz):
(>100−500 Hz):
(>500 Hz−1 kHz):
Power Factor:
Voltage Magnitude:
Current Magnitude (Low Range)
Fundamental:
Harmonics 2−49:
Current Magnitude (High Range)
Fundamental:
Harmonics 2−49:
Agilent 6811B/6812B/6813B
0.05% + 10 mA
0.05% + 15 mA
0.05% + 30 mA
0.05% + 1.5 mA
0.05% + 8 mA
0.05% + 25 mA
0.05% + 150 mA
0.03% + 150 mA
0.1% + 1.5 VA + 1.2 mVA/V
0.1% + 2 VA + 1.2 mVA/V
0.1% + 6 VA + 1.2 mVA/V
0.1% + 1.5 VA + 12 mVA/V
0.1% + 2 VA + 12 mVA/V
0.1% + 6 VA + 12 mVA/V
0.1% + 0.3 W + 1.2 mW/V
0.1% + 1.2 W + 1.2 mW/V
0.1% + 2.5 W + 1.2 mW/V
0.1% + 0.3 W + 12 mW/V
0.1% + 1.2 W + 12 mW/V
0.1% + 2.5 W + 12 mW/V
0.01
0.03% + 100 mV + 0.2%/kHz
0.03% + 1.5 mA
0.03% + 1 mA + 0.2%/kHz
0.05% + 5 mA
0.03% + 3 mA + 0.2%/kHz