Agilent Technologies 6811B User Manual page 79

Ac power solutions
Hide thumbs Also See for 6811B:
Table of Contents

Advertisement

1110
PRINT TABXY(15,9),"3. Connect the 3458A across the current shunt"
1120
PRINT TABXY(15,11),"4. Set the 3458A to AC rms VOLTS"
1130
PRINT TABXY(15,13),"5. Turn on the AC Source"
1140
!
1150
DISP "Press CONT to begin Current Program and Measurement calibration"
1160
PAUSE
1170
CLEAR SCREEN
1180
PRINT TABXY(22,5),"CALIBRATING CURRENT POGRAMMING"
1190
PRINT TABXY(20,7),"There are 2 points to be calibrated"
1191
OUTPUT @Ac;"CAL:STATE ON"
1200
OUTPUT @Ac;"CAL:CURR:AC"
1210
OUTPUT @Ac;"CAL:LEV P1"
1220
WAIT 10
1230
INPUT "Enter AC rms current ( DVM reading divided by shunt resistance )",Ai_p1
1240
PRINT TABXY(25,9),"Point 1 entered"
1250
OUTPUT @Ac;"CAL:DATA";Ai_p1
1260
OUTPUT @Ac;"CAL:LEV P2"
1270
WAIT 10
1280
INPUT "Enter AC rms current ( DVM reading divided by shunt resistance )",Ai_p2
1290
PRINT TABXY(25,11),"Point 2 entered"
1300
OUTPUT @Ac;"CAL:DATA";Ai_p2
1310
WAIT 10
1320
CLEAR SCREEN
1330
PRINT TABXY(22,5),"CALIBRATING CURRENT MEASUREMENT"
1340
PRINT TABXY(20,7),"There are 2 points to be calibrated"
1350
OUTPUT @Ac;"CAL:CURR:MEAS"
1360
OUTPUT @Ac;"CAL:LEV P1"
1370
WAIT 10
1380
INPUT "Enter AC rms current ( DVM reading divided by shunt resistance )",Am_p1
1390
PRINT TABXY(25,9),"Point 1 entered"
1400
OUTPUT @Ac;"CAL:DATA";Am_p1
1410
OUTPUT @Ac;"CAL:LEV P2"
1420
WAIT 10
1430
INPUT "Enter AC rms current ( DVM reading divided by shunt resistance )",Am_p2
1440
PRINT TABXY(25,11),"Point 2 entered"
1450
OUTPUT @Ac;"CAL:DATA";Am_p2
1460
WAIT 10
1470
OUTPUT @Ac;"CAL:SAVE"
1480
OUTPUT @Ac;"CAL:STATE OFF"
1490
CLEAR SCREEN
1500
!
1510
!
1520
PRINT TABXY(15,5),"1. Turn off the AC Source"
1530
PRINT TABXY(15,7),"2. Disconnect all equipment from the AC Source"
1540
PRINT TABXY(15,9),"3. Connect the 1 ohm impedance resistor, see fig.B-1"
1550
PRINT TABXY(15,11),"4. Turn on the AC Source"
1560
!
1570
DISP "Press CONT to begin Output Impedance calibration"
1580
PAUSE
1590
PRINT TABXY(15,10),"CALIBRATING OUTPUT IMPEDANCE"
1600
OUTPUT @Ac;"CAL:STATE ON"
1610
OUTPUT @Ac;"CAL:IMP"
1620
PRINT TABXY(30,15),"WAIT"
1630
WAIT 30
1640
OUTPUT @Ac;"CAL:SAVE"
1650
OUTPUT @Ac;"CAL:STATE OFF"
1660
CLEAR SCREEN
1670
PRINT TABXY(25,10),"CALIBRATION COMPLETE"
1680
END
Figure B-2. Calibration Program Listing (Sheet 3 of 3)
Verification and Calibration - B
79