B - Verification and Calibration
560
OUTPUT @Ac;"CAL:LEV P1"
570
WAIT 10
580
INPUT "Enter DC voltage reading from DVM",Dc_p1
590
PRINT TABXY(25,9),"Point 1 entered"
600
OUTPUT @Ac;"CAL:DATA";Dc_p1
610
OUTPUT @Ac;"CAL:LEV P2"
620
WAIT 10
630
INPUT "Enter DC voltage reading from DVM",Dc_p2
640
PRINT TABXY(25,11),"Point 2 entered"
650
OUTPUT @Ac;"CAL:DATA";Dc_p2
660
OUTPUT @Ac;"CAL:LEV P3"
670
WAIT 10
680
INPUT "Enter DC voltage reading from DVM",Dc_p3
690
PRINT TABXY(25,13),"Point 3 entered"
700
OUTPUT @Ac;"CAL:DATA";Dc_p3
710
WAIT 3
720
CLEAR SCREEN
730
PRINT TABXY(5,10),"1. Connect the 3458A to the rear terminals via the ratio transformer"
740
PRINT TABXY(18,12),"2. Set the 3458A to AC VOLTS"
750
!
760
DISP "Press CONT to begin AC PROGRAMMING and MEASUREMENT calibration"
770
PAUSE
780
CLEAR SCREEN
790
PRINT TABXY(18,5),"CALIBRATING AC POGRAMMING and MEASUREMENT"
800
PRINT TABXY(20,7),"There are 4 points to be calibrated"
810
OUTPUT @Ac;"CAL:VOLT:AC"
820
OUTPUT @Ac;"CAL:LEV P1"
830
WAIT 10
840
INPUT "Enter AC rms ( transformer ratio * DVM reading)",Ac_p1
850
PRINT TABXY(25,9),"Point 1 entered"
860
OUTPUT @Ac;"CAL:DATA";Ac_p1
870
OUTPUT @Ac;"CAL:LEV P2"
880
WAIT 10
890
INPUT "Enter AC rms ( transformer ratio * DVM reading)",Ac_p2
900
PRINT TABXY(25,11),"Point 2 entered"
910
OUTPUT @Ac;"CAL:DATA";Ac_p2
920
OUTPUT @Ac;"CAL:LEV P3"
930
WAIT 10
940
INPUT "Enter AC rms ( transformer ratio * DVM reading)",Ac_p3
950
PRINT TABXY(25,13),"Point 3 entered"
960
OUTPUT @Ac;"CAL:DATA";Ac_p3
970
OUTPUT @Ac;"CAL:LEV P4"
980
WAIT 3
990
INPUT "Enter AC rms ( transformer ratio * DVM reading)",Ac_p4
1000
PRINT TABXY(25,15),"Point 4 entered"
1010
OUTPUT @Ac;"CAL:DATA";Ac_p4
1020
WAIT 10
1030
CLEAR SCREEN
1040
PRINT TABXY(15,10),"CALIBRATING OVERVOLTAGE PROTECTION"
1050
OUTPUT @Ac;"CAL:VOLT:PROT"
1060
PRINT TABXY(30,15),"WAIT"
1070
WAIT 30
1071
OUTPUT @Ac;"CAL:SAVE"
1072
OUTPUT @Ac;"CAL:STATE OFF"
1080
CLEAR SCREEN
1090
PRINT TABXY(15,5),"1. Turn off the AC Source"
1100
PRINT TABXY(15,7),"2. Connect the current shunt and 20 ohm load resistor, see fig.B-1"
78