Agilent Technologies 6811B User Manual page 77

Ac power solutions
Hide thumbs Also See for 6811B:
Table of Contents

Advertisement

10
!
20
! AC Source calibration program
30
!
40
ASSIGN @Ac TO 705
50
!
60
PRINT TABXY(5,5),"This program will calibrate the 6811B/12B/13B AC Power Solutions."
70
PRINT TABXY(5,7),"Equipment requirements are:
80
PRINT TABXY(37,8),"0.01 ohm <200ppm Current Shunt"
90
PRINT TABXY(37,9),"20 ohm >1800 watt power resistor for all models"
100
PRINT TABXY(37,10),"1 ohm >100 watt impedance resistor"
110
PRINT TABXY(37,11),"30:1 <50ppm Ratio Transformer"
120
PRINT TABXY(3,13),"Ratio Transformer is required to when calibrating to MIL-STD-45662A. If the"
130
PRINT TABXY(2,14),"ratio transformer is not used the measurement uncertainty must be recalculated."
140
!
150
DISP "Press CONT to continue"
160
PAUSE
170
CLEAR SCREEN
180
PRINT TABXY(15,5),"1. Turn the AC Source off"
190
PRINT TABXY(15,7),"2. Disconnect all loads"
200
PRINT TABXY(15,9),"3. Connect the 3458A to the rear terminal block"
210
PRINT TABXY(15,11),"4. Set the 3458A to DC VOLTS"
220
PRINT TABXY(15,13),"5. Turn on the AC Source"
230
!
240
DISP "Press CONT to begin DC OFSET and DC GAIN calibration"
250
PAUSE
260
CLEAR SCREEN
270
PRINT TABXY(25,5),"CALIBRATING VOLTAGE OFFSET"
280
PRINT TABXY(20,7),"There are 4 points to be calibrated"
290
OUTPUT @Ac;"CAL:STATE ON"
300
OUTPUT @Ac;"CAL:VOLT:OFFS"
310
OUTPUT @Ac;"CAL:LEV P1"
320
WAIT 10
330
INPUT "Enter DC offset voltage reading from DVM",Off_p1
340
PRINT TABXY(25,9),"Point 1 entered"
350
OUTPUT @Ac;"CAL:DATA";Off_p1
360
OUTPUT @Ac;"CAL:LEV P2"
370
WAIT 10
380
INPUT "Enter DC offset voltage reading from DVM",Off_p2
390
PRINT TABXY(25,11),"Point 2 entered"
400
OUTPUT @Ac;"CAL:DATA";Off_p2
410
OUTPUT @Ac;"CAL:LEV P3"
420
WAIT 10
430
INPUT "Enter DC offset voltage reading from DVM",Off_p3
440
PRINT TABXY(25,13),"Point 3 entered"
450
OUTPUT @Ac;"CAL:DATA";Off_p3
460
OUTPUT @Ac;"CAL:LEV P4"
470
WAIT 10
480
INPUT "Enter DC offset voltage reading from DVM",Off_p4
490
PRINT TABXY(25,15),"Point 4 entered"
500
OUTPUT @Ac;"CAL:DATA";Off_p4
510
WAIT 3
520
CLEAR SCREEN
530
PRINT TABXY(25,5),"CALIBRATING DC VOLTAGE GAIN"
540
PRINT TABXY(20,7),"There are 3 points to be calibrated"
550
OUTPUT @Ac;"CAL:VOLT:DC"
Figure B-2. Calibration Program Listing (Sheet 1 of 3)
Verification and Calibration - B
Rev B.00.00
Agilent3458A or equivalent DVM"
77