Model Agilent 6811B
Test Description
300 Vrms accuracy at 45 Hz
Front Panel Measurement
300 Vrms accuracy at 400 Hz
Front Panel Measurement
300 Vrms accuracy at 1 kHz
Front Panel Measurement
425 Vdc accuracy
Front Panel Measurement
−425 Vdc accuracy
Front Panel Measurement
Low Range
High Range
(Iout = calculated output current @ 60Hz)
Model Agilent ______________________
Test Description
300 Vrms accuracy at 45 Hz
Front Panel Measurement
300 Vrms accuracy at 400 Hz
Front Panel Measurement
300 Vrms accuracy at 1 kHz
Front Panel Measurement
Table B-2. Agilent 6811B Verification Test Record
Report No.____________ Date_____________
Minimum
Specification
Voltage Programming and Measurement Accuracy
299.250 V
Vrms −190 mV
298.200 V
Vrms −190 mV
296.700 V
Vrms −400 mV
DC Programming and Measurement Accuracy
424.075 V
Vdc −0.277 V
−424.075 V
Vdc −0.277 V
RMS Current Measurement Accuracy
Iout −0.0039 A
Iout −0.0095 A
Table B-3. Agilent 6812B Verification Test Record
Minimum
Specification
Voltage Programming and Measurement Accuracy
299.250 V
Vrms −190 mV
298.200 V
Vrms −190 mV
296.700 V
Vrms −400 mV
Verification and Calibration - B
Recorded Results
_______V
_______V
_______V
_______V
_______V
_______V
_______V
_______V
_______V
_______V
_______A
_______A
Report No.____________
Recorded Results
_______V
_______V
_______V
_______V
_______V
_______V
Maximum
Specification
300.750 V
Vrms +190 mV
301.800 V
Vrms +190 mV
303.300 V
Vrms +400 mV
425.925 V
Vrms +0.277 V
−425.925 V
Vrms +0.277 V
Iout +0.0039 A
Iout +0.0095 A
Date_____________
Maximum
Specification
300.750 V
Vrms +190 mV
301.800 V
Vrms +190 mV
303.300 V
Vrms +400 mV
69