To Control Switching Matrix - Agilent Technologies B1500A Training Manual

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Module 2
Getting Started

To Control Switching Matrix

Step 1: Click Switching Matrix button to open the Switching Matrix Operation Panel.
Step 2: Enter the switch setup name.
Step 3: Define the switch setup by using the matrix on the operation panel window. Label can be set
to input ports and output channels.
Step 4: Click the Apply Switch Setup button to send the switch setup to the B2200/E5250 switching
matrix.
Note:
Before clicking this button, connect the B2200/E5250 to the B1500 via GPIB, and set the
configuration. To set the configuration, see Module 12.
The switch setups you create or modify MUST be saved in a preset group.
To save the setup, click the Save button.
To recall the setup, click the Recall button.
You can organize the setups by using the Preset List button.
To select the available preset group, use the field below the Preset List button. In the above example,
the DemoSW group is selected and listed in the Preset List list.
To rename, duplicate, or delete the setups, and to add, rename, duplicate, or delete the preset group,
use the function of the Preset List button menu or the Organize Preset Group dialog box shown in the
next page.
Read Switch Setup: Read switch setup from B2200/E5250 via GP-IB and show it on the operation
panel window.
Note: Standard revision of EasyEXPERT cannot support the E5250. The E5250 control is optional
function of B1540A-002/B1541A-002.
Note: EasyEXPERT supports the E5250A installed with the E5252A matrix card.