Remote Programming - Source Output Commands; Table 3-2 Dc Output Commands - Keithley 6220 DC User Manual

6220 dc current source 6221 ac and dc current source
Table of Contents

Advertisement

Model 6220/6221 User's Manual
Remote programming – source output commands
Table 3-2
ming example to output DC current is also provided.
Table 3-2
DC output commands
Command
CLEar
CURRent:RANGe <n>
CURRent:RANGe:AUTO <b>
CURRent <n>
CURRent:COMPliance <NRf>
CURRent:FILTer <b>
OUTPut:RESPonse <name>
OUTPut <b>
SourceMeter
1. To select a fixed source range, specify the current output value that is going to
be sourced. The Model 622x will go to the lowest range that can source that
value. For example, If you are going to source 25mA, let <n> = 25e-3. The
100mA range will be selected.
2. Selecting a fixed source range disables autorange.
3. The output must be off in order change the output response. Sending this
command while the output is on will generate error -220 Execution Error.
4. OUTP OFF turns the output off, but does change the set output level.
SOUR:CLE sets the output level to zero and then turns the output off.
lists the commands to configure and control the DC output. A program-
Turns output off and sets output level to zero.
Sets current source range (amps).
<n> = -105e-3 to 105e-3
Enables or disables source autorange.
<b> = ON or OFF
Sets DC current source output level (amps).
<n> = -105e-3 to 105e-3
Sets voltage compliance (volts).
<NRf> = 0.1 to 105
Enables or disables the output analog filter.
<b> = ON or OFF
Select fast or slow output response speed for 6221.
<name> = FAST or SLOW
Turn output on or off.
<b> = ON or OFF
Sets output to zero, then turns the output off
Return to
DC Current Source Operation 3-15
Description
3, 4
Section 3 topics
Default
1, 2
100e-3
OFF
0.0
10.0
OFF
FAST
OFF
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

6221 dc6221 ac

Table of Contents