Getting Started

Default settings

The Model 622x can be restored to one of seven setup configurations: five user-
saved setups, PRESET (bench defaults), and *RST (bus defaults). As shipped
from the factory, the Model 622x powers up to the PRESET settings. PRESET
settings provide a general purpose setup for front panel operation, while the *RST
settings do the same for remote operation.
NOTE User setups cannot be saved or recalled while Wave,
See Section 1 of the Reference Manual for more information on "Default settings"
including a table that lists all the default settings.

Save and restore setups

Save setup:
Press SAVE > Select SAVE > Set memory location (0 to 4) > Press ENTER
Restore user setup:
Press SETUP > Select USER > Set memory location (0 to 4) > Press ENTER
Restore PRESET defaults:
Press SETUP > Select PRESET
Restore *RST defaults:
Press SETUP > Select *RST
Remote programming
*SAV <NRf>
*RCL <NRf>
SYSTem:PRESet
*RST
Sweep, Delta, Pulse Delta, or Differential Conductance is
armed or running. Attempting to do so will generate error
+413 Not allowed with mode armed.
A custom sweep cannot be saved as a user setup. At-
tempting to do so will generate error +528 Cannot save
CUSTOM sweep setup.
Source preset values are not saved as part of a user set-
up.
' Save present setup in memory.
'
Recall saved user setup from memory.
<NRf> = 0, 1, 2, 3, or 4
'
' Restore PRESET default setup.
' Restore *RST default setup.