Buffer

The Model 622x has a buffer (data store) to store readings and related data ele-
ments for one to 65,536 buffer readings. Buffer readings are readings from the
Model 2182/2182A that were processed by the Model 622x as delta readings
(Delta, Pulse Delta, or Differential Conductance). Also stored are statistics for the
readings. These include maximum, minimum, average mean, standard deviation,
and peak-to-peak.
For front panel operation, the buffer is always active (enabled). Post-math delta
readings are automatically stored in the buffer.
Remote programming provides more control of the buffer:
NOTE For remote programming, only the buffer recall commands

Buffer characteristics

Buffer size
The buffer size automatically sets to accommodate the number of delta readings
to be performed. For example, if the test is configured to perform 1,000 delta mea-
surements, the buffer size will be set to 1000.
If the delta test is configured to perform an infinite number of delta measurements,
the buffer will be set to maximum size (65,536 readings). After the 65,536th read-
ing, the delta test continues but readings stop being stored.
Buffer data elements
Data elements are stored along with each delta reading. The data elements for
front panel operation are shown in
timestamp, reading units, and the current source output value.
For remote programming, the returned reading string for a buffer read command
includes the reading and the data elements in the following order:
Reading with Units, Timestamp, Current Source, Average Voltage, Compliance
State, and Reading Number.
The buffer can be disabled.
When enabled, buffer feed can be pre-math delta readings or post-math
delta readings.
The timestamp can be set for the absolute format or the delta format.
are documented in this section. See Section 6 of the Refer-
ence Manual for details on all buffer commands.