Table 5-4 Differential Conductance Commands - Keithley 6220 DC User Manual

6220 dc current source 6221 ac and dc current source
Table of Contents

Advertisement

Model 6220/6221 User's Manual
Table 5-4
Differential Conductance commands
SOURce[1]:DCONductance:NVPResent?
SOURce[1]:DCONductance:STARt <NRf> Sets start value (amps).
SOURce[1]:DCONductance:STEP <NRf>
SOURce[1]:DCONductance:STOP <NRf>
SOURce[1]:DCONductance:DELTa <NRf> Sets delta value (amps).
SOURce[1]:DCONductance:DELay <NRf> Sets delay (seconds).
SOURCE[1]:DCONductance:CABort <b>
SOURce[1]:DCONductance:ARM
SOURce[1]:DCONductance:ARM?
SOURce[1]:DCONductance:NVZero?
1. Use the :NVPR? command to determine if a suitable Model 2182/2182A with
the correct firmware revision is properly connected to the RS-232 port. This
query command can be used for the system configuration shown in
Figure
The RS-232 cannot be the selected interface for the Model 6221. If it is, this
query will generate error -221 Settings Conflict.
2. See
tings (:STARt, :STEP, :STOP, :DELTa, :DELay, and :CABort).
3. After setting up Differential Conductance using the above setup commands,
the :ARM command arms Differential Conductance. During the arming pro-
cess, the Model 622x communicates with the Model 2182/2182A.
When armed, Differential Conductance will start when the Model 6221 is trig-
gered. Differential Conductance can be "un-armed" by sending the following
command: SOURce:SWEep:ABORt.
Command
5-2B on
page
5-5.
"Configuration
settings" on
Return to
Delta, Pulse Delta, and Differential Conductance 5-43
Description
Queries if 2182/2182A is con-
1
nected.
1 = yes, 0 = no
<NRf> = -105e-3 to 105e-3
Sets step size (amps).
<NRf> = 0 to 105e-3
Sets stop value (amps).
<NRf> = -105e-3 to 105e-3
<NRf> = 0 to 105e-3
<NRf> = 1e-3 to 9999.999
Enable or disable Compliance
2
Abort.
<b> = 0 or OFF, 1 or ON
Arms Differential Conductance.
Queries Diff Cond arm.
1 = armed, 0 = not armed.
Queries V-zero value acquired from
4
2182/2182A.
page 5-37
for more information on these set-
Section 5 topics
Default
2
-10e-6
2
1e-6
2
10e-6
2
1e-6
2
0.002
0
3
3

Advertisement

Table of Contents
loading

This manual is also suitable for:

6221 dc6221 ac

Table of Contents