Welch Allyn 1500 Directions For Use Manual page 83

Table of Contents

Advertisement

Directions for use
Approved catheters and injectate sensor type
Catheters
The monitor will work with any blood temperature sensor that is Edwards-compatible.
Injectate Sensor
The injectate sensor must be a Baxter/Edwards/Abbott "in-line" thermistor.
Note
Preparing for cardiac output measurement
Note
1.
Press the Setup button
enabled (yes).
2. Connect the catheter to the CO connector on the side panel.
3. Setup the catheter according to the instructions given by the manufacturer.
The monitor does not support bath probe or arrow flow-through sensor.
The CO is only displayed when the CO option is enabled. If the CO measurement
field is not displayed, ensure that it is enabled in the parameter settings.
, enter the menu Setup/Parameters and activate CO
Monitoring and Measurements
79

Advertisement

Table of Contents
loading

Table of Contents