Welch Allyn 1500 Directions For Use Manual page 130

Table of Contents

Advertisement

126
Technical data
Capnography
Module
Measuring method
CO
CO
Curve Resolution
etCO
CO
Respiration Rate range
Respiration Rate
Accuracy
Flow rate
Waveform sampling
Initialization Time
System Response Time
Calibration Interval
Note
Mini Medi CO
Non dispersive Infrared Spectroscopy
units
mmHg or kPa
2
, etCO
, fiCO
, range
0 to 99 mmHg (CO
2
2
2
0.1 mmHg
, inCO
Resolution
1 mmHg
2
2
Accuracy
0 to 38 mmHg: ± 2 mmHg
2
39 to 99 mmHg: ± 5 % of reading and 0.08 % for every 1 mmHg above 38
mmHg
0 to 150 Resp/min
0 to 70: ± 1 Resp/min
71 to 120: ± 2 Resp/min
121 to 150: ± 3 Resp/min
50 ml/min, (42.5 < flow < 65) flow measured by volume
20 samples/s
40 s (typical)
5.6 s (typical combined response time)
Initially calibrate after 1,200 operating hours, then once a year or
after 4,000 operating hours, whichever comes first. The initial
calibration should not occur before 720 hours of use. If the initial
calibration is done before 720 hours of use, the module will reset
to require its next calibration after 1200 hours, instead of after
4000 hours.
The capnography component of this product is covered by one or more of the following US patents:
6,428,483; 6,997,880; 5,300,859; 6,437,316; 7,488,229; 7,726,954 and their foreign equivalents.
Additional patent applications pending.
Welch Allyn 1500 Patient Monitor