Monitoring and Measurements
Pacemaker monitoring
3- and 5-lead cables for pacemaker patients
The following illustration shows the electrode placement with a 5-lead patient cable for
optimum results for patients with an implanted pacemaker.
With a 3-lead patient cable, only R, L and F are connected.
IEC
Black (N)
Red (R)
Yellow (L)
Green (F)
White (C)
WARNING Patients with a pacemaker must be observed continuously because
the heart rate from the pacemaker might still be registered in case of a cardiac
arrest or some arrhythmias. See specification
disclosure of the pacemaker pulse rejection capability of this monitor.
WARNING Pacemaker monitoring is not possible with ECG cables that have un-
shielded lead wires. Ensure that only shielded lead wire ECG cables are used
when monitoring patients that have a pacemaker.
WARNING Welch Allyn recommends using an SpO
ECG measurement and to set the alarm range for the peripheral pulse (PP) in the
range of the heart rate (HR), or to set the HR source in the SpO
(see
"SpO
settings"
on page 69.)
2
White (C)
Red (R)
Black (N)
AHA
Green (RL)
White (RA)
Black (LA)
Red (LL)
Brown (V)
"Technical data"
Yellow (L)
Green (F)
Welch Allyn 1500 Patient Monitor
on page 117 for
sensor in addition to the
2
menu to SpO
2
2