Directions for use
NIBP settings
The parameter settings are selected with the trim knob, described previously (see
"Settings via a parameter field"
The default settings are in bold.
Parameter
NIBP Interval
Format
Initial Inflation Pressure
NIBPs (systolic blood pressure)
upper/lower limit
NIBPd (diastolic blood pressure)
upper/lower limit
NIBPm (mean pressure) upper/
lower limit
on page
31).
Description
Off, 3, 5, 10, 15, 30, 60
Sys/Dia or Mean. This defines the main measurement to be
displayed, that is the larger measurement displayed in the NIBP box.
The secondary measurement is displayed smaller by the side of the
defined measurement. Note that the mean measurement is shown in
brackets.
Adult: 160
Pediatric: 120
Neonate: 90
Range: 30 - 300
Adult 220/75
Pediatric: 145/75
Neonate: 100/50
Range: 20 - 235
Adult 110/35
Pediatric: 100/35
Neonate: 70/30
Range: 20 - 255
Adult 120/50
Pediatric: 110/50
Neonate: 80/35
Monitoring and Measurements
65