Directions for use
Masimo Module
Sensors
Amplifier
Sampling frequency
Display update interval
Signal IQ waveform
Perfusion Index numeric
(PI)
Fast SAT
Display ranges
Averaging Time
Sensitivity
Accuracy
Masimo
Fully isolated, defibrillation protected >5kV
62.5 Hz
1 second
A waveform that indicates pulse detection confidence. Values range from
0 to 127 where 0 is low confidence and 127 is high confidence.
A numeric provided that indicates perfusion. Perfusion is measured in %
and ranges from 0.02% to 20%.
A mode that enables rapid tracking of arterial oxygen saturation changes
by minimizing the averaging.
Saturation (SpO
)
0 to 100%
2
PR
25 to 240 beats per minute
2 / 4 / 8 / 10 / 12 / 14 / 16 / 18
Normal / Max / APOD
SpO
, no motion
60 to 80 ± 3%, adults / pediatrics
2
70 to 100 ± 2%, adults / pediatrics, ± 3% neonates
SpO
, motion
70 to 100 ± 3%, adults / pediatrics / neonates
2
SpO
, low perfusion
70 to 100 ± 2%, adults / pediatrics / neonates
2
PR, no motion
25 to 240 ± 3 bpm, adult s/ pediatrics / neonates
PR, motion
25 to 240 ± 5 bpm, adults / pediatrics / neonates
PR, low perfusion
25 to 240 ± 5 bpm, adults / pediatrics / neonates
®
®
LNCS
SET
125
Technical data