Settings
Restore user defaults
The saved user settings are enabled. The user-defined settings are saved in the Setup
System menu (see
Administrator
The administrator and service screens provide system information and option settings.
The Administrator screens are entered from the settings menu:
Administrator sub-menus can only be accessed with a password. The passwords are
divided into clinical, service and factory passwords and only clinical settings are available
for the user. The other menu options on this menu are for Welch Allyn service personnel.
Options are enabled in the service menu; please contact Welch Allyn. The passwords for
the alarms and system sub-menus are as follows:
Setup > Administrator > Alarms: 49, 48, 46 (Clinical Password).
Setup> Administrator > System 49, 48, 46 (Clinical Password).
No password is needed to view the Configuration. Other sub-menus are intended for
service personnel only.
The Administrator menu is as follows:
Sub menu
Configuration
Communications
Alarms
(password
protected - see
above)
"Saving the user-defined settings as default"
Setup Menu > Administrator
Parameter
Alarm Silence Time
Alarm Suspend Time
Can disable HR / PR
Alarms
Alarm Delay
Second Speaker Time
Description
Monitor data such as serial number, software version etc. This
is provided for information only and no settings can be made.
This provides the communication settings for service
personnel only.
1, 1.5 or 2 minutes. Time for which an audible alarm is silenced.
1, 1.5, or 2 minutes. Time for which all alarms are suspended.
Enabled/Disabled. The menu item can prevent users from
being able to turn off HR or PR alarms. When the setting is
"Disabled", a user cannot turn off the HR or PR alarms.
On/Off. If the alarm validation is enabled, alarm limits must be
exceeded for at least 6 seconds for an alarm to be issued.
0 to 3 minutes (2 minutes). Time after which the secondary
speaker is enabled. For lethal and high level alarms the
secondary speaker is activated in 30 seconds.