Monitoring and Measurements
IBP alarm defaults
Label / Parameter
P1, P2, P3, P4 SYS
P1, P2, P3, P4 DIA
P1, P2, P3, P4 Mean
ART SYS
ART DIA
ART Mean
CVP SYS
CVP DIA
CVP Mean
PA SYS
PA DIA
PA Mean
ICP SYS
ICP DIA
ICP Mean
RA SYS
RA DIA
RA Mean
LA SYS
LA DIA
LA Mean
UA SYS
UA DIA
UA Mean
UV SYS
UV DIA
UV Mean
Adult
Adult
Lower Limit
Upper Limit
6
14
-4
6
0
10
75
220
35
110
50
120
6
14
-4
6
0
10
10
34
0
16
0
20
6
14
-4
6
0
10
6
14
-4
6
0
10
6
14
-4
6
0
10
Welch Allyn 1500 Patient Monitor
Pediatric
Pediatric
Lower Limit
Upper Limit
2
10
-4
2
0
4
75
145
35
100
50
110
2
10
-4
2
0
4
10
34
0
16
0
20
2
10
-4
2
0
4
2
10
-4
2
0
4
2
10
-4
2
0
4
Neonate
Neonate
Lower Limit
Upper Limit
2
10
-4
2
0
4
50
100
30
70
35
80
2
10
-4
2
0
4
10
34
0
16
0
20
2
10
-4
2
0
4
2
10
-4
2
0
4
2
10
-4
2
0
4
50
100
30
70
35
80
2
10
-4
2
0
4