Unpacking and installation
Fig. 14
14
5
6.2

Assembling the handwheel

Before the instrument can be used, the
handwheel has to be assembled to the main
body. All required parts and tools are con-
tained in the shipping crate.
The feather key (5) is loosely placed in the handwheel
shaft (1) and secured during transport with a tie-rap.
• Remove the tie-rap.
Attention - Make sure the feather key does not fall
out!
• Place the handwheel (2) onto the handwheel shaft
(1) as shown.
• Tighten the screw located in the center hole of the
handwheel with an Allen key, size 5 (3).
• Remove the lidding foil from the autoadhesive cover
(4) and place it onto the handwheel.
To disassemble, proceed in reverse order.
6.3

Site requirements

Do not operate the instrument in rooms with
explosion hazard!
To ensure trouble-free function, a minimum
distance of at least 10 cm to walls and other
objects must be kept on all sides of the instru-
ment.
The installation site must fulfill the following require-
ments:
- stable, vibration-free laboratory bench,
- virtually vibration-free floor,
- Easy and comfortable access to the handwheel,
- Room temperature constantly between +10 °C and
+ 35 °C,
- relative air humidity not exceeding 80 %.