Setting Up The Instrument; Connecting To Mains - Leica RM2145 Instruction Manual

Rotary microtome
Hide thumbs Also See for RM2145:
Table of Contents

Advertisement

3
Fig. 15
Leica RM 2145 – Rotary Microtome
2
230 V
7.1

Connecting to mains

7.1.1 Checking the voltage selector setting
1
The voltage selector is located in the voltage selector
housing (3) above the mains switch (2) on the rear of
the instrument. The actual voltage setting can be seen
in a small indication window (1) in the voltage selector
housing.
• Remove the adhesive tape.
• Check if the voltage selector setting that shows in
the window (1) is the appropriate one for your local
mains supply.
If the setting is correct, please proceed to section 7.1.3
of this manual.
If the voltage shown in the indication window does not
correspond to your local mains supply, the voltage se-
lector setting MUST be changed prior to connecting
the instrument to the mains supply.
7.

Setting up the instrument

Caution: The voltage selector is preset in the
factory.
Before connecting the instrument to the mains
power, please check whether the factory set-
ting is identical with the local mains power
in your lab!
The mains cable inlet is sealed with an adhe-
sive tape that indicates the factory setting the
instrument comes with.
Severe damage may be caused to the instru-
ment when operating it at an incorrect volt-
age setting.
15