Comparing the Control Solution Test Results

The test result of each control solution should be within the
range printed on the label of the test strip vial or on the test
strip box. Repeat the control solution test if the test result
falls outside of the range. Out of range results may occur in
following situations:
Situations
• When the control solutions are not used
with their corresponding test strips,
• When the control solution bottle was
not shaken well,
• When the meter, test strip, or the
control solution were exposed to high
or low temperatures,
• When the first drop of the control
solution was not discarded or the tip of
the bottle was not wiped clean,
• When the meter is not functioning
properly.
• When the control solution is past the
expiration date printed on the bottle,
• When the control solution is past its
discard date,
• When the control solution is
contaminated.
If results continue to fall outside the range, the test strip and
meter may not be working properly. Do not use your system
and contact i-SENS sales representative.
34 www.i-sens.com
Do This
Repeat the
control solution
test by referring
to the notes
on page 31.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.

Using the Lancing Device

You will need a lancing device in order to collect a blood
sample.
You may use the lancing device that is included in the CareSens
Dual System or any other medically approved lancing device.
Adjustable Tip
Lancet Holder
Load Confirmation
Lancet
Window
• The lancing device is for use by a single user only and should
not be shared with anyone.
• Use a soft cloth or tissue to wipe the lancing device. If
necessary, a small amount of alcohol on a soft cloth or tissue
may be used.
Caution:
To avoid infection when drawing a sample, do not use a
lancet more than once, and:
• Do not use a lancet that has been used by others.
• Always use a new sterile lancet.
• Keep the lancing device clean.
Note:
Repeated puncturing at the same sample site may cause pain
or skin calluses (thick hard skin). Choose a different
site each time you test.
Ejection System
Sliding Barrel
Release Button
www.i-sens.com 35