29.6.27 List of available operating modes

Command: OMI CR LF
Possible replies:
OMI˽ CR LF
N° designation CR LF
N° designation CR LF
OK CR LF
˽
OMI
I CR LF
n
1
2
3
4
5
6
7
8
9
10 - Totalization
11 - Tolerance weighing
12 - Statistics
13 - Pipette calibration

29.6.28 Set operating mode

Command: OMS CR LF
Possible replies:
OMS˽OK CR LF
˽
OMS
E CR LF
˽
OMS
I CR LF

29.6.29 Query current operating mode

Command: OMG CR LF
Possible replies:
OMG˽n˽OK CR LF
˽
OMG
I CR LF
n
Number of operating mode, see chap. 29.6.27
168
Command accepted, available operating modes are shown
Command valid but cannot be executed at the moment
Operating mode
- Weighing
- Parts counting
- Percent determination
- Dosing
- Make up recipes
- Animal weighing
- Density determination "solid matter"
- Density determination "liquids"
- Maximum value function
Command valid
No parameter or incorrect format
Command valid but cannot be executed at the moment
Command accepted, number of current operating mode will be
shown
Command valid but cannot be executed at the moment
AEJ-C/AES-C/PLJ-C-BA-e-1621