21 Animal weighing
The Animal weighing application facilitates weighing of unstable loads or weighing
under unstable ambient conditions. The weighing balance calculates the mean value
of several weighing results obtained over a set time.
Two different start / reset modes are available.
Manual (via keystroke, see chap. 21.1)
or
Automatic (automatic start after exceeding a minimum weight, see chap.
21.2).
21.1 Animal weighing by manual start
Preferences:
Applications can be adapted to your requirements by creating application-related
settings. The settings available depend on the currently enabled application.
A part of these settings is global, that is, they are – apart from a few exceptions –
identical to the weighing application (See chap 9.3 „Settings <P2.1 WEIGHING>".
The description below merely deals with settings deviating from the default settings.
Menu item
<P2.9.7 AUTOSTART>
<P2.9.7 MEAN. Z.>
<P2.9.4> SHORTCUTS
Functions keys
F1; F2; F3; F4
AEJ-C/AES-C/PLJ-C-BA-e-1621
Adjustment Description
<NO>
The measuring cycle is started manually,
that is, placing of load must be confirmed
by keystroke.
selectable
Duration after which the weighing balance
1 – 90 s
calculates the mean value of several
weighing results.
<START>
Assign key with "Measuring cycle
manual start" function
111