KERN AEJ-C Operating Manual page 134

Analysis and precision balances
Hide thumbs Also See for AEJ-C:
Table of Contents

Advertisement

Printout example KERN YKB-01N
-------------------Pipette calibration---------------------
Meas. No.
Date
Time
Temp.
Humidity
Pressure
-------------------------Tested volume 1000 µl------------
1
2
3
4
5
6
7
8
9
10
Average Volume [Va]
Systematic errorr [ES]
Random error [Sr]
---------------------- Tested volume 5000 µl------------
1
2
3
4
5
6
7
8
9
10
Average Volume [Va]
Systematic errorr [ES]
Random error [Sr]
--------------------- Tested volume 10000 µl-------------
---
1
2
3
4
5
6
7
8
9
10
Average Volume [Va]
Systematic error [ES]
Random error [Sr]
-------------------------------------------------------------------
Signature
..................................................................
134
10
05.01.2016
11:34:17
20.0°C
50 %
1017hPa
1003 µl
993 µl
1013 µl
1003 µl
993 µl
1003 µl
1013 µl
1053 µl
1003 µl
1023 µl
1010 µl
Arithmetic average
1.03 %
Systematic measurement deviation
17.7 µl
Standard deviation
4966 µl
4966 µl
4966 µl
4986 µl
4976 µl
4966µl
4966 µl
4976 µl
4976 µl
4976 µl
4972 µl
Arithmetic average
0.56 %
Systematic measurement deviation
7.0 µl
Standard deviation
10033 µl
10033 µl
10033 µl
10033 µl
10043 µl
10043 µl
10043 µl
10043 µl
10043 µl
10043 µl
10039 µl
Arithmetic average
0.39 %
Systematic measurement deviation
5.2 µl
Standard deviation
AEJ-C/AES-C/PLJ-C-BA-e-1621