National Instruments RMX-4120 User Manual page 64

Rmx programmable power supplies
Hide thumbs Also See for RMX-4120:
Table of Contents

Advertisement

CF21 Command Language Parameter
Selects the command language that is used in remote control messages. For the LAN interface
to function, this parameter must be set to "SCPi."
This parameter is applied when you turn the POWER switch off and on.
Value
SCPi
LGCy
CF22 Emulation Parameter During Remote Control
Selects the emulation that is used during remote control. If you have set the command language
to legacy languages, you can only select "GEn" or "PAG."
This parameter is applied when you turn the POWER switch off and on.
Value
nonE
5700
GEn
PAG
dCS
CF23 SCPI Communication Error Display Parameter
Selects whether to display communication errors. The errors are only displayed when you are
using the SCPI language.
Value
oFF
on
The SCPi language will be used (factory default setting).
Legacy languages will be used.
Remote control is performed using the RMX programmable
power supply's commands made by National Instruments
(factory default setting).
Remote control is performed using the N5700/N8700 commands
made by Agilent Technologies.
Remote control is performed using the Genesys series commands
made by TDKLambda.
Remote control is performed using the PAG series commands
made by Kikusui.
Remote control is performed using the DCS series commands
made by Sorensen.
SCPI communication errors are not displayed (factory default
setting).
SCPI communication errors are displayed.
RMX Programmable Power Supplies User Manual
Parameter Description
Parameter Description
Parameter Description