Page 2
RMX-410x Rackmount Programmable DC Power Supplies 200W/400W/800W Built-in USB, RS-232, RS-485, LAN Interface USER MANUAL This Manual Covers Models: RMX-4101 RMX-4102 RMX-4104 NATIONAL INSTRUMENTS...
Table of Contents CHAPTER 1: GENERAL INFORMATION 1.1 User Manual Content ......................... 13 1.2 Introduction........................... 13 1.2.1 General Description ......................13 1.2.3 Features and Options ...................... 13 1.2.4 Multiple Output Power System ..................14 1.2.6 Analog Voltage Programming and Monitoring ............. 14 1.2.7 Parallel Operation ......................
Page 5
3.9.10 Multiple Load Connection with Distribution Terminals ........40 3.9.11 Grounding Outputs ......................41 3.10 Local and Remote Sensing ..................... 41 3.10.1 Sense Wiring ........................41 3.10.2 Local Sensing ........................42 3.10.3 Remote Sensing........................ 42 3.10.4 J2 Sense Connector Technical Information ............43 3.11 Repackaging for Shipment .....................
Page 6
5.5 Parallel Operation ........................61 5.5.1 Introduction ......................... 61 5.5.2 Basic Parallel Operation ....................61 5.5.2.1 Master Unit Set Up ......................61 5.5.2.2 Slave Unit Set Up ......................62 5.5.2.3 Setting Over Voltage Protection................62 5.5.2.4 Setting Foldback Protection ..................62 5.5.2.5 Connection to Load .......................
Page 7
7.7.6 Backspace ..........................81 7.7.7 Error Messages ........................81 7.8 GEN Command Set Description....................82 7.8.1 General guides ........................82 7.8.2 Command Set Categories ....................82 7.8.3 Identification Commands ....................82 7.8.4 Initialization Commands ....................83 7.8.5 Output Commands ......................83 7.8.6 Global Output Commands .....................
Page 8
8.7.2 Waveform Example ......................120 CHAPTER 9: STATUS, FAULT AND SRQ REGISTERS 9.1 General ............................121 9.2 Power Supply Status Structure ....................122 9.3 Condition Registers ........................122 9.3.1 Fault Register ........................122 9.3.2 Status Register ........................123 9.4 Conditional, Enable and Event Registers ................123 9.4.1 Conditional Registers.
Page 9
NATIONAL INSTRUMENTS Electromagnetic Compatibility Guidelines This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic environment.
Page 10
EU Customers: At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, visit ni.com/...
Page 11
RMX-410x series units are not authorized for use as critical component in nuclear control systems, life support systems or equipment for use in hazardous environments without the express written approval of the National Instruments. OVERVOLTAGE CATEGORY AND ENVIRONMENTAL CONDITIONS The RMX-410x series units have been evaluated to Overvoltage category II.
Page 12
NATIONAL INSTRUMENTS PARTS SUBSTITUTIONS & MODIFICATIONS Parts substitutions and modifications are by authorized NI service personnel only. For repairs or modifications, the instrument must be returned to NI service facility. AC INPUT Do not connect RMX-410x series unit to mains supply exceeding the input voltage and frequency rating.
Page 13
NATIONAL INSTRUMENTS WARNING: There is a potential shock hazard when using a power supply with an output voltage greater than 60VDC. Do not turn ON power supply when output voltage is above 60VDC without output bus-bars and output connectors protection assembled. Turn OFF power supply or disconnect power supply from AC mains before making or changing any rear panel connection.
NATIONAL INSTRUMENTS CHAPTER 1: GENERAL INFORMATION 1.1 User Manual Content This user's manual contains the operating instructions, installation instructions and specifications of the RMX-410x Series 200W, 400W, and 800W power supply series. The instructions refer to the standard power supplies, including the built-in USB and RS232/485 serial communication. For information related to operation with the LAN, refer to User Manual for power supply LAN.
NATIONAL INSTRUMENTS 1.2.4 Multiple Output Power System The RMX-410x Series power supplies series can be configured into a programmable power system of up to 31 units using the built-in LAN, USB or RS232/RS485 communication port in the power supply and the RS485 linking cable provided with each power supply. In a LAN system, each power supply can be controlled using the LAN controller (installed).
NATIONAL INSTRUMENTS CAUTION: Observe all torque guidelines within this manual. Over torque may damage unit or accessories. Such damage is not covered under manufacturers warranty. 1.3 Accessories 1.3.1 General Accessories are delivered with the power supply or separately upon ordering, The list below shows the possible accessories and ordering numbers.
2.1 RMX-4101 Series Specifications RMX-4101 20-10 36-6 60-3.5 100-2 MODEL 1. Rated output voltage(*1) 2. Rated output current (*2) 3. Rated output power CONSTANT VOLTAGE MODE 20-10 36-6 60-3.5 100-2 1. Max. Line regulation (*6) 0.01% of rated output voltage+2mV 2.
Page 18
ANALOG PROGRAMMING AND MONITORING 1. Vout voltage programming 0~100%, 0~5V or 0~10V, user selectable. Accuracy and linearity: +/-0.5% of rated Vout. 2. Iout voltage programming (*13) 0~100%, 0~5V or 0~10V, user selectable. Accuracy and linearity: +/-1% of rated Iout. 3. Vout resistor programming 0~100%, 0~5/10Kohm full scale, user selectable.
Page 19
RMX-4101 Series Specifications RMX-4101 INPUT CHARACTERISTICS 20-10 36-6 60-3.5 100-2 1. Input voltage/freq. (*3) 85~265Vac continuous, 47~63Hz, single phase 2. Maximum Input current 100/200VAC (*4) (*18) 2.62/1.29 2.76/1.37 2.69/1.33 2.55/1.26 3. Power Factor (Typ) >0.99 at 100Vac, >0.98 at 200Vac,100% load 4.
Page 20
NOTES: *1: Minimum voltage is guaranteed to maximum 0.1% of rated output voltage. *2: Minimum current is guaranteed to maximum 0.2% of rated output current. *3: For cases where conformance to various safety standards (UL, IEC, etc…) is required, to be described as 100-240Vac (50/60Hz). *4: Ta=25°C with rated output power.
2.2 RMX-4102 Series Specifications RMX-4102 MODEL 20-20 36-12 60-7 100-4 1. Rated output voltage(*1) 2. Rated output current (*2) 3. Rated output power 20-20 36-12 60-7 100-4 CONSTANT VOLTAGE MODE 1. Max. Line regulation (*6) 0.01% of rated output voltage+2mV 2.
Page 22
ANALOG PROGRAMMING AND MONITORING 1. Vout voltage programming 0~100%, 0~5V or 0~10V, user selectable. Accuracy and linearity: +/-0.5% of rated Vout. 2. Iout voltage programming (*13) 0~100%, 0~5V or 0~10V, user selectable. Accuracy and linearity: +/-1% of rated Iout. 3. Vout resistor programming 0~100%, 0~5/10Kohm full scale, user selectable.
Page 23
RMX-4102 Series Specifications RMX-4102 INPUT CHARACTERISTICS 20-20 36-12 60-7 100-4 1. Input voltage/freq. (*3) 85~265Vac continuous, 47~63Hz, single phase 2. Maximum Input current 100/200VAC (*4) (*18) 4.98/2.45 5.25/2.57 5.10/2.50 4.80/2.37 3. Power Factor (Typ) 0.99 at 100/200Vac, 100% load 4. Efficiency (Typ) 100/200VAC (*4) (*18) 81/83 83/85 83/85...
Page 24
NOTES: *1: Minimum voltage is guaranteed to maximum 0.1% of rated output voltage. *2: Minimum current is guaranteed to maximum 0.2% of rated output current. *3: For cases where conformance to various safety standards (UL, IEC, etc…) is required, to be described as 100-240Vac (50/60Hz). *4: Ta=25°C with rated output power.
2.3 RMX-4104 Series Specifications RMX-4104 MODEL 20-40 36-24 60-14 100-8 1. Rated output voltage(*1) Vin ≥ 100Vac, Ta ≤ 50°C 2. Rated output 85Vac ≤ Vin < 100Vac, Ta ≤ 40°C current (*2)(*21) 85Vac ≤ Vin < 100Vac, 40°C < Ta ≤ 50°C 12.5 Vin ≥...
Page 26
ANALOG PROGRAMMING AND MONITORING 1. Vout voltage programming 0~100%, 0~5V or 0~10V, user selectable. Accuracy and linearity: +/-0.5% of rated Vout. 2. Iout voltage programming (*13) 0~100%, 0~5V or 0~10V, user selectable. Accuracy and linearity: +/-1% of rated Iout. 3. Vout resistor programming 0~100%, 0~5/10Kohm full scale, user selectable.
Page 27
RMX-4104 Series Specifications RMX-4104 20-40 36-24 60-14 100-8 INPUT CHARACTERISTICS 1. Input voltage/freq. (*3) 85~265Vac continuous, 47~63Hz, single phase 2. Maximum Input current 100/200VAC (*4) 9.65/4.75 10.30/5.10 10.00/4.95 9.50/4.7 3. Power Factor (Typ) 0.99 at 100Vac, 100% load / 0.98 at 200Vac, 100% load 4.
Page 28
*8: Measured with JEITA RC-9131A (1:1) probe. *9: From 10% to 90% or 90% to 10% of Rated Output Voltage, with rated resistive load. *10: From 90% to 10% of Rated Output Voltage. *11: For load voltage change, equal to the unit voltage rating, constant input voltage. *13: The Constant Current programming, readback and monitoring accuracy do not include the warm-up and Load regulation thermal drift.
NATIONAL NATIONAL INSTRUMENTS INSTRUMENTS CHAPTER 3: INSTALLATION 3.1 General This chapter contains instructions for initial inspection, preparation for use and repackaging for shipment. Connection to PC, setting the communication port and linking RMX-410x power supplies are described in Chapter 7. NOTE: RMX-410x power supplies generate magnetic fields which might affect the operation of other instruments.
NATIONAL INSTRUMENTS 3.5 Location, Mounting and Cooling This power supply is fan cooled. The air intake is at the front panel and the exhaust is at the rear panel. Upon installation allow cooling air to reach the front panel ventilation inlets. Allow minimum 10 cm (4”) of unrestricted air space at the front and the rear of the unit.
NATIONAL INSTRUMENTS 3.7.2 AC Input Cord Refer to section 1.3.4 for details of the AC input cords recommended. WARNING: The AC input cord connector is the disconnect device of the power supply. The connector must be readily identifiable and accessible to the user. The AC input cord must be no longer than 3m. 3.8 Turn-On Checkout Procedure 3.8.1 General The following procedure ensures that the power supply is operational and may be used as a basic...
NATIONAL INSTRUMENTS 3.8.3 Constant Voltage Check Turn on the output by pressing OUTPUT button so the OUTPUT LED illuminates. Observe the power supply Voltage display and rotate the Voltage encoder. Ensure that the output voltage varies while the Voltage encoder is rotated. The minimum control range is from zero to the maximum rated output for the power supply model.
NATIONAL INSTRUMENTS 3.8.7 Foldback Check WARNING: There is a potential shock hazard when checking a power supply with output voltage greater than 60VDC. Observe proper safety procedures during the checking. Refer to Section 5.3.4 for explanation of the FOLD function prior to performing the procedure below. 1.
Page 37
NATIONAL INSTRUMENTS Wire size Resistivity Maximum length in feet to limit Ohm/1000ft voltage drop to 1V or less 2.526 1.589 0.9994 12.5 0.6285 0.3953 0.2486 0.1564 1200 Table 3-2: Maximum wire length for 1V drop on lead (in feet) Cross sect. Resistivity Maximum length in meters to limit area...
NATIONAL INSTRUMENTS 3.9.3 Wire Termination The wires should be properly terminated with terminals securely attached. DO NOT use non terminated wires for load connection at the power supply. CAUTION: When local sensing, a short from +LS or +S to -V or -S or -LS, will cause damage to the power supply. Reversing the sense wires might cause damage to the power supply in local and remote sensing.
Page 39
NATIONAL INSTRUMENTS CAUTION: Ensure that the load wiring mounting hardware does not short the output terminals. Heavy connecting cables must have some form of strain relief to prevent loosening the connections or bending the bus-bars. 20V to 100V Models Refer to Fig.3-4 for connection of the load wires to the power supply bus-bars and to Fig.3-5 for mounting the bus-bars shield to the chassis.
NATIONAL INSTRUMENTS 3.9.7 Connecting Single Loads, Local Sensing (default) Fig.3-6 shows recommended load and sensing connections for a single load. The local sense lines shown are default connections at the rear panel J2 sense connector. Local sensing is suitable for applications where load regulation is less critical.
NATIONAL INSTRUMENTS 3.9.9 Connecting Multiple Loads, Radial Distribution Method Fig.3-8 shows multiple loads connected to one supply. Each load should be connected to the power supply’s output terminals using separate pairs of wires. It is recommended that each pair of wires will be as short as possible and twisted or shielded to minimize noise pick-up and radiation.
NATIONAL INSTRUMENTS 3.9.11 Grounding Outputs Either the positive or negative output terminals can be grounded. To avoid noise problems caused by common-mode current flowing from the load to ground, it is recommended to ground the output terminal as close as possible to the power supply chassis ground. Always use two wires to connect the load to the power supply regardless of how the system is grounded.
NATIONAL INSTRUMENTS 3.10.2 Local Sensing The power supply is shipped with the rear panel J2 sense connector wired for local sensing of the output voltage. Refer to Table 3-4 for J2 terminals assignment. With local sensing, the output voltage regulation is made at the output terminals. This method does not compensate for voltage drop on the load wires, therefore it is recommended only for low load current applications or where the load regulation is less critical.
NATIONAL INSTRUMENTS CAUTION: When using shielded sense wires, ground the shield in one place only. The location can be the power supply chassis or one of the output terminals. Use remote sense where the load regulation at the load end is critical. In remote sense, the power supply will compensate for voltage drop on the load wires.
NATIONAL INSTRUMENTS CHAPTER 4: FRONT/REAR PANEL CONTROLS AND CONNECTORS 4.1 Introduction The RMX-410x Power Supply series has a full set of controls, indicators and connectors that allow the user to set up and operate the unit. Before starting to operate the unit, please read the following sections for an explanation of the functions, controls and connector terminals.
Page 46
NATIONAL INSTRUMENTS Control/Indicator Description Section AC Power Switch AC ON/OFF control 4 digit 7-segment LED display. Normally displays the output current. Current display In preview mode, the display indicates the program setting of output current. 4 digit 7-segment LED display. Normally displays the output voltage. Voltage Display In preview mode, the display indicates the program setting of output voltage.
NATIONAL INSTRUMENTS 4.3 Rear Panel Connectors Refer to Fig.4-2 and Table 4-2 for description of the Rear Panel connectors. Fig. 4-2: Rear panel connections Connection Description Section AC Input Connector IEC320-16 TYPE CONNECTOR DC output bus-bar Bus-bars for 10V to 100V models. Use M6 or 1/4” screws. Analog Control and Connector for remote analog interface.
Page 48
NATIONAL INSTRUMENTS WARNING: Terminals 7, 9 and 12 of J1 are connected internally to the negative sense(-S) potential of the power supply. Do not attempt to bias any of these terminals relative to the negative sense. Use the Isolated Programming interface option to allow control from a programming source at a different potential relative to the power supply negative.
NATIONAL INSTRUMENTS 4.3.1 J1 Connector Terminal and Function Control and monitoring signals are referenced to the negative sense potential (-S). Connector Technical Information: Connector type: IPL1-106-01-S-D-RA-K, SAMTEC • Receptacle type: IPD1-06-D-K, SAMTEC • • Contact pins: CC79R-2024-01-L, SAMTEC • Hand tool: CAT-HT-179-2024-11, SAMTEC •...
NATIONAL INSTRUMENTS 4.3.2 J3 Connector Terminal and Function Control and monitoring signals are isolated from the power supply output. Connector Technical Information • Connector type: IPL1-104-01-S-D-RA-K, SAMTEC • Receptacle type: IPD1-04-D-K, SAMTEC • Contact pins: CC79R-2024-01-L, SAMTEC • Hand tool: CAT-HT-179-2024-11, SAMTEC •...
NATIONAL INSTRUMENTS 4.4 Front Panel Display Messages Table 4-5 shows the various messages that will be shown on the display in different operating modes. Display Text Text Description Display Text Text Description Abor ABORT ONCE ONCE ADDRESS OT ATO AUTO (RESTART) O...
NATIONAL INSTRUMENTS 4.5 Navigating the Main Menu 4.5.1 Introduction The Main Menu consists of three levels: Subsystem, Function and Parameter. To enter the Menu press the Menu button. The Menu LED illuminates and the display shows the Subsystem Menu. Navigate by rotating the Voltage encoder to scroll through the Subsystem list (first level). Repeat these actions to navigate the Functions list (second level).
NATIONAL INSTRUMENTS 4.5.2 Exiting the Main Menu There are three ways to exit from Main Menu: 1. Press MENU button twice. MENU LED turns OFF. Display shows present status of power supply. 2. Press and hold MENU button 3sec. MENU LED turns OFF. Display shows present status of power supply.
NATIONAL INSTRUMENTS 4.6.2 Exiting the Communication Menu There are three ways to exit from REM menu: 1. Press REM button twice. REM LED turns OFF. Display shows present status of power supply. 2. Press and hold REM button 3sec. REM LED turns OFF. Display shows present status of power supply. 3.
NATIONAL INSTRUMENTS CHAPTER 5: LOCAL OPERATION 5.1 Introduction This Chapter describes the operating modes that do not require programming and monitoring the power supply via its communication interface. USB or RS232/RS485 or by remote analog signals. Ensure that the REM LED on the front panel is Off, (indicating Local mode). If the REM LED is On, press the front panel REM button to change the operating mode to local.
NATIONAL INSTRUMENTS 5.2.3 Automatic Crossover When the power supply operates in Constant Voltage mode, while the load current is increased to greater than the current limit setting, the power supply will automatically switch to Constant Current mode. If the load is decreased to less than the current limit setting, the power supply will automatically switch back to Constant Voltage mode.
NATIONAL INSTRUMENTS 5.3 Alarms and Protective Functions 5.3.1 Introduction There are several conditions that cause alarm (RED LED blinks). All alarms affect the output. When an alarm occurs, the respective fault will appear on the display and the alarm LED illuminates. It is possible that more than one fault (alarm) may be triggered but only the first will be shown on the display.
NATIONAL INSTRUMENTS 5.3.2.2 Resetting the OVP Circuit To reset the OVP circuit after activation: 1. Reduce the power supply Output Voltage setting below the OVP set level. 2. Ensure that the load and the sense wiring is connected properly. 3. Four methods to reset the OVP circuit. •...
NATIONAL INSTRUMENTS 5.3.4.1 Setting the Foldback Protection The Foldback can be set when the power supply output is Enabled (On) or Disabled (Off ). 1. Press PROT button. PROT (GREEN) LED illuminates. The ”O” message appears on the Voltage display. 2.
NATIONAL INSTRUMENTS 5.4 Series Operation Power supplies of the same model can be connected in series to obtain increased output voltage. Split connection of the power supplies gives positive and negative output voltage. WARNING: When power supplies are connected in series, and the load or one of the output terminals is grounded, no point may be at a greater potential of +/- 100VDC from ground.
NATIONAL INSTRUMENTS 5.4.2 Series Connection for Positive and Negative Output Voltage In this mode, two units are configured as positive and negative output. Set the current limit of each power supply to the maximum that the load can handle without damage. It is recommended that diodes be connected in parallel with each unit output to prevent reverse voltage during start-up or in case one of the units shuts down.
NATIONAL INSTRUMENTS 5.5 Parallel Operation 5.5.1 Introduction Up to six units of the same Voltage and Current rating can be connected in parallel to provide up to six times the output current capability. One of the units operates as a master and the remaining units are slaves.
NATIONAL INSTRUMENTS 5.5.2.2 Slave Unit Set Up When Slave mode is selected the power supply enters Current programming mode via external Voltage. Voltage and Current programming setting values are set to 105% of range. During operation the slave units operate as a controlled current source following the master output current. It is recommended that the power system is designed so that each unit supplies up to 95% of its current rating.
Page 64
NATIONAL INSTRUMENTS Fig.5-5: Parallel operation with remote sensing CAUTION: Make sure that the connection between - Vo terminals is reliable to avoid disconnection during operation. Disconnection may cause damage to the power supply. NOTE: With local sensing it is important to minimize the wire length and resistance. Also the positive and negative wire resistance should be as close as possible to each other to achieve current balance between power supplies.
NATIONAL INSTRUMENTS 5.5.3 Advanced Parallel Operation In Advanced Parallel operation the master unit displays the total current of all units connected in Parallel. The slave units display ”ON SE”. The master and slave units operate in a Daisy-Chain connection configuration. For further details about Daisy-chain connection refer to section 5.6. In the Advanced Parallel mode, the total current is programmed and reported by the master unit.
NATIONAL INSTRUMENTS 5.6 Daisy-Chain Connection It is possible to configure a multiple power supply system to shut down all units when a fault condition occurs in one of the units. When the fault is removed, the system recovers according to a preset state: Safe start mode or Automatic restart.
NATIONAL INSTRUMENTS 5.7.1 External Shut Off Function SO signal serves as Output Shut Off . It is an optically isolated signal from the power supply output. Connection to the signal is made via pin J3-5 (Shut Off) and pin J3-7 (IFC_COM). The SO pin accepts a 4V to 15V signal or Open-Short contact to disable or enable the power supply output.
NATIONAL INSTRUMENTS Front Panel ILC Power Supply ILC Input Display Alarm LED Setting Output OFF - Default Open or Short Voltage/Current Open Blinking Short Voltage/Current Table 5-5: Interlock functions and settings CAUTION: To prevent possible damage to the unit, do not connect any of the Enable /Disable inputs to the positive or negative output potential.
NATIONAL INSTRUMENTS 5.7.4 Power Supply OK Signal PS_OK signal indicates fault condition in the power supply. It is a TTL signal output at J3-2, referenced to IFC_COM at J3-7 (Isolated Interface Common). When a fault condition occurs, PS_OK level is low, with maximum sink current of 1mA. When no fault condition occurs, PS_OK level is high with maximum source current of 2mA.
NATIONAL INSTRUMENTS 5.9 Parameter Setting Memory Power Supply has four memory configuration modes: Subsystem Function Parameter Display Display Display Description Level Level Level SAVE SAE 1…4 1...4 Save setting in non violate memory RECALL REC 1…4 1...4 Recall setting in non violate memory Memory MEMo es...
NATIONAL INSTRUMENTS 5.9.4 Save <1..4> This command saves the present state of the power supply to a specified location in memory (refer to Table 5-7). Up to 4 states can be stored. Storage locations 1 through 4 are in nonvolatile memory.
Page 72
NATIONAL INSTRUMENTS Parameter Factory Default Resetting Last setting Save & Recall Output Status Voltage Set-point Current Set-point Fold Back mode Over Voltage Protection Under Voltage Level/ OFF (UVL) OFF (UVL) Protection mode Under Voltage Level/ Protection level Auto Start Mode SAFE SAFE Control pin 1...
NATIONAL INSTRUMENTS CHAPTER 6: REMOTE ANALOG PROGRAMMING 6.1 Introduction The Rear Panel connector J1 allows the user to program the power supply output voltage and current limit with an analog device. J1 also provides monitoring signals for output voltage and output current.
NATIONAL INSTRUMENTS 6.4 Remote Voltage Programming of Output Voltage and Current Remote Programming settings are as follows: 1. For Voltage Analog Programming wiring refer to Fig.6-1. 2. Short pins J1-1 to J1-7. 3. Press MENU button. MENU (GREEN) LED illuminates. ”Set” message appears on Voltage display. 4.
NATIONAL INSTRUMENTS 6.5 Remote Resistor Programming of Output Voltage and Output Current For resistive programming, internal current sources, for output voltage and/or output current control, supply 1mA current through external programming resistors connected between J1-6 and J1-12 and between J1-5 and J1-1, J1-7 & J1-11. The voltage across the programming resistors is used as a programming voltage for the power supply.
NATIONAL INSTRUMENTS NOTES: In Remote analog mode: the output voltage cannot be set by the Voltage encoder. The output voltage limit is set to 5% over the model-rated maximum value. The output Current limit is set by the Current encoder to 5% over the model-rated maximum value. The power supply will operate within the extended range, however it is not recommended to operate the power supply over its voltage and current rating and performance is not guaranteed.
NATIONAL INSTRUMENTS CHAPTER 7: Serial RS232/RS485 and USB Interface 7.1 Introduction This chapter describes the set-up, operation, commands and communication protocol of RMX- 410x power supplies via serial communication interfaces: RS232, RS485 or USB. 7.2 Configuration Function Level Display Parameter Level Display 45 Interface...
NATIONAL INSTRUMENTS 7.2.4 Baud Rate Setting Seven optional rates are possible: 1200, 2400, 4800, 9600, 19200, 38400, 57600. 1. Press REM button. The REM LED illuminates. ”I Nt” message appears on the Voltage display. 2. Rotate Voltage encoder until Voltage display shows ”b d”. 3.
NATIONAL INSTRUMENTS 7.3 Rear Panel RS232/485 Connector The RS232/485 interface is accessible through the Rear panel RS232/485 IN and RS485 OUT connectors. The connectors are 8 contact RJ-45. The IN and OUT connectors are used to connect power supplies in a RS232 or RS485 chain to a controller. Refer to Fig.7-1 for IN/OUT connectors. Shield (Connector enclosure) NC 8...
NATIONAL INSTRUMENTS 7.4 Connectig Power Supply To RS232 Or RS485 BUS Connect rear panel IN connector to the controller RS232 or RS485 port using a suitable shielded cable. Refer to Figures 7-3 and 7-4 for available RS232 and RS485 cables. Socket DB-9 CONNECTOR 8 PIN CONNECTOR...
NATIONAL INSTRUMENTS 7.5 Rear Panel USB Connector A standard USB Series B device connector is located on Rear panel for USB control. Refer to Fig.7-5 and Table 7-2. Designator Description VBUS +5 VDC Data - Data + Interface com Fig.7-5: USB Connector Table 7-2: USB connector pin out 7.5.1 USB Getting Started USB Cable (*)
NATIONAL INSTRUMENTS NOTES: It is recommended when using ten or more power supplies in Daisy-chain system to connect 120Ω resistive termination at the last unit’s RS-485 out connector 120Ω, 0.5W between TXD+ and TXD-. 120Ω, 0.5W between RXD+ and RXD-. L=0.5m typ.
NATIONAL INSTRUMENTS Error Code Description Returned when program voltage (PV) is programmed above acceptable range. Example: PV above 95% of OVP setting. Returned when programming output voltage below UVL setting. Returned when OVP is programmed below acceptable range. Example: OVP value is less than 105% of voltage setting. Returned when UVL value is programmed above the programmed output voltage.
NATIONAL INSTRUMENTS 7.8.4 Initialization Commands Command Description ADR n ADR is followed by address which can be 1 to 31 and is used to access the power supply . Clear status. Sets FEVE and SEVE registers to zero. Reset command. Brings the power supply to a safe and known state: Output voltage: zero, Remote: non-latched remote, Output current: zero,...
Page 85
NATIONAL INSTRUMENTS Command Description Turns the output to ON or OFF. Recover from Safe-Start, OVP or FLD fault. OUT 1 (or OUT OUT n ON)-Turn On. Returns the output On/Off status string. OUT? ON- output on. OFF- output off. Sets the Foldback protection to ON or OFF. FLD 1 (or FOLD ON) - Arms the Foldback protection.
NATIONAL INSTRUMENTS 7.8.6 Global Output Commands General Global commands can be received by all power supplies connected to the BUS, without individual address commands. All power supplies will execute the command immediately. There is no acknowledgment back to the PC when using global commands. A delay must be set of 20msec after each global command.
Page 87
NATIONAL INSTRUMENTS Model Minimum (A) Maximum (A) Model Minimum (A) Maximum (A) 20-10 00.0000 10.0000 20-20 00.0000 20.0000 36-6 0.00000 6.00000 36-12 00.0000 12.0000 60-3.5 0.00000 3.50000 60-7 0.00000 7.00000 100-2 0.00000 2.00000 100-4 0.00000 4.00000 Table 7-6: RMX-4101 models Current programming range Table 7-7: RMX-4102 models Current programming range and communication response format and communication response format...
NATIONAL INSTRUMENTS 7.9 Serial Communication Test Set-Up Basic set-up to test serial communication operation. PC with Windows Hyper Terminal, private edition, software installed, RMX-410x power 1. Equipment: supply, RS232 cable. 2. PC set-up: 2.1 Open Hyper Terminal.......New Connection. 2.2 Enter a name 2.3 Connect to........Direct to Com1 or Com 2 2.4 Configure port properties: Bits per second ..9600...
NATIONAL INSTRUMENTS 7.10.4 Checksum The user may optionally add a checksum to the end of the command. The checksum is ”$” followed by two hex characters. If a command or a query has checksum, the response will also have one. There is no CR between the command string and the ”$”...
NATIONAL INSTRUMENTS 7.10.8 Data Formats Data Formats Description Digits with an implied decimal point assumed at the right of the least-significant digit. <NR1> Examples: 256 <NR2> Digits with an explicit decimal point. Example: .0253 <NR3> Digits with an explicit decimal point and an exponent. Example: 2.73E+2 <NRf>...
Page 92
Serial number, Firmware level’ . Meaning and Type Identification System Interface Query Syntax *IDN? Field Information National Instruments Manufacturer Model Returned Parameters <Vrating>-<Irating> Model ratings 25B1234 Serial number, typically 7 alpha-numeric characters 3.0-C1 Revisions, <Main firmware>–<LAN/IEEE firmware> Example National Instruments,RMX20-30,25B1234, 1.0-C1...
Page 93
NATIONAL INSTRUMENTS *OPC Operation Complete command. Sets the Operation Complete bit in the Standard Event Status Register if all commands and queries are completed. Meaning and Type Operation Complete Device Status Command Syntax *OPC Parameters None *OPC? Operation Complete query. Returns ASCII ‘1’ as soon as all commands and queries are completed. Meaning and Type Operation Complete Device Status...
Page 94
NATIONAL INSTRUMENTS *RCL n Restores the power supply to a state previously stored in memory by *SAV command. Refer to Table 5-7. Command Syntax *RCL <NR1> Parameters 1 to 4 Exmple *RCL 3 *RST This command resets the power supply to a defined state as shown in Table 5-7. *RST also forces an ABORt command.
Page 95
NATIONAL INSTRUMENTS *STB? Status Byte query. Returns the contents of the Status Byte Register. Meaning and Type Status Byte Device Status Query Syntax *STB? Returned Parameters <NR1> (Register binary value) Bit Position Condition OPER QUES (RQS) Bit Weight ESB = Event status byte summary; MAV = Message available MSS = Master status summary;...
NATIONAL INSTRUMENTS 7.12 SCPI Subsystem Commands Subsystem commands are specific to power supply functions. They can be a single command or a group of commands. Groups are comprised of commands that extend one or more levels below the root. Commands followed by a question mark (?) take only the query form. Except as noted in the syntax descriptions, all other commands take both the command and query form.
Page 97
NATIONAL INSTRUMENTS OUTPut:PROTection:FOLDback Foldback mode is used to disable the output when a transition is made between the operation modes. The power supply will turn off the output after a specified delay if the power supply makes transition into CV mode or into CC mode. This feature is particularly useful for protecting current or voltage sensitive loads.
Page 98
NATIONAL INSTRUMENTS OUTPut:TTLTrg:MODE Sets the operation of the Trigger Out signal to either OFF, Function Strobe or Trigger mode. Programming Mode NONE, FIX: • In TRIG mode, trigger is generated when output status changes. • In Function Strobe mode, an output pulse is generated automatically any time an output parameter such as output, voltage or current is programmed.
NATIONAL INSTRUMENTS 7.12.2 Instrument Subsystem The subsystem programs one or more power supplies. <NR1> from 1 to 31. INSTrument:COUPle SCPI Command Syntax INSTrument:COUPle <CRD> GEN Command Syntax None Parameters ALL|NONE INSTrument:NSELect SCPI Command Syntax INSTrument:NSELect <NRf> GEN Command Syntax ADR n (single address) Examples INSTrument:NSELect 6 Query Syntax...
Page 100
NATIONAL INSTRUMENTS NOTE: VOLT:MODE LIST and WAVE is an implied ABORT command. WAVE mode cannot be programmed simultaneously for both Voltage and Current. Only the last sent command can be accepted as WAVE. Previous mode reverts to NONE. VOLTage:PROTection:LEVel Sets the OVP level. The OVP setting range is given in Table 7-9. The number of characters after OVP is up to 12.
NATIONAL INSTRUMENTS VOLTage:PROTection:LOW Sets the under voltage protection (UVP) level of the power supply. SCPI Command Syntax [SOURce]:VOLTage:PROTection:LOW <NRf+> GEN Command Syntax UVL < NRf+> , UVP < NRf+> Default Suffix *RST Value Examples VOLT:PROT:LOW 2.5 VOLT:PROT:LOW MAX [SOURce]:VOLTage:PROTection:LOW [:LEVel]? Query Syntax VOLT:PROT:LOW? MIN;...
NATIONAL INSTRUMENTS CURRent:MODE This command selects FIX, LIST, WAVE subsystems control over the power supply output current. SCPI Command Syntax [SOURce]:CURRent:MODE <CRD> GEN Command Syntax None Parameters NONE | FIXed | LIST | WAVE *RST Value NONE Examples CURR:MODE LIST CURR:MODE FIX Query Syntax [SOURce]:CURRent:MODE? Returned Parameters...
NATIONAL INSTRUMENTS 7.12.7 INITiate Subsystem INITiate Enables the trigger subsystem. If a trigger circuit is not enabled, all trigger commands are ignored. SCPI Command Syntax INITiate[:IMMediate] GEN Command Syntax None Parameters None Examples INIT:IMM Query Syntax None INITiate:CONTinuous • INIT:CONT 0 - Enables the trigger subsystem only for a single trigger action. The subsystem must be enabled prior to each subsequent trigger action.
Page 105
NATIONAL INSTRUMENTS LIST:CURRent Specifies the output current points in a list. The current points are given in the command parameters, which are separated by commas. SCPI Command Syntax [SOURce]:LIST:CURRent <NRf+> {,<NRf+>} Default Suffix Examples LIST:CURR 2.5,3.0,3.5 LIST:CURR MAX,2.5,MIN up to 12 parameters Query Syntax LIST:CURRent? Returned Parameters...
NATIONAL INSTRUMENTS LIST:VOLTage Specifies the output voltage points in a list. The voltage points are given in the command parameters, which are separated by commas. SCPI Command Syntax [SOURce]:LIST:VOLTage <NRf+> {,<NRf+>} Default Suffix Examples LIST:VOLT 2.0,2.5,3.0 LIST:VOLT MAX,2.5,MIN up to 12 parameters Query Syntax LIST:VOLT? Returned Parameters...
Page 107
NATIONAL INSTRUMENTS STATus:OPERation:ENABle Sets the value of the Enable register. This register is a mask for enabling specific bits from the Condition register to the Event register. SCPI Command Syntax STATus:OPERation:ENABle <NRf> GEN Command Syntax SENA nnnn, SENA? Parameters 0…7FFF Returned Parameters <NR1>...
NATIONAL INSTRUMENTS 7.12.10 SYSTem Subsystem SYSTem:ERRor:ENABle Enables Error messages. SCPI Command Syntax SYSTem:ERRor:ENABle GEN Command Syntax None Parameters None Returned Parameters None SYSTem:ERRor? Returns the next error number and corresponding error message in the power supply error queue. Works as FIFO. When no error exists 0, ”No error” is returned. SCPI Command Syntax SYSTem:ERRor? GEN Command Syntax...
NATIONAL INSTRUMENTS SYSTem:DATE? SCPI Command Syntax SYSTem:DATE? GEN Command Syntax DATE? Query Syntax SYSTem:DATE? Returned Parameters <CRD> yyyy/mm/dd SYSTem:PON:TIME? Time measured from first power On. SCPI Command Syntax SYSTem:PON:TIME? Parameters Minute Example 6534 Returned Parameters <NR1> 7.12.11 TRIGger Subsystem NOTE: The Trigger subsystem must be enabled from the Initiate subsystem or no triggering action will occur.
NATIONAL INSTRUMENTS TRIGger:SOURce Selects the power supply input trigger source as follows: • BUS (*TRG & TRIG) and Front Panel • EXT Mainframe backplane Trigger IN PIN SCPI Command Syntax TRIGger[:STARt]:SOURce <CRD> GEN Command Syntax None Parameters BUS | EXTernal *RST Value EXTernal Examples...
NATIONAL INSTRUMENTS WAVE:STEP • WAVE:STEP AUTO - When triggered, creates waveforms consecutively, until the wave is completed. • WAVE:STEP ONCE - When triggered, it executes one step from the list. SCPI Command Syntax [SOURce]WAVE:STEP <CRD> Example WAVE:STEP AUTO Query Syntax [SOURce]:WAVE:STEP? Returned Parameters AUTO | ONCE...
NATIONAL INSTRUMENTS Global Commands List Result From Command Name Description SCPI Format GEN Format Supply Enable all Outputs Output On to last V GLOBal:OUTPut:STATe 1|ON GOUT 1 None & I levels Disable all Outputs Turn outputs Off to GLOBal:OUTPut:STATe 0|OFF GOUT 0 None zero V &...
Page 113
NATIONAL INSTRUMENTS Subsystem Commands SCPI Command Description GEN Command ABORt Aborts the triggered action <NC> DISPlay [:WINDow]:STATe <bool> Display ON/OFF <NC> [:WINDow]:FLASh <bool> Display Flash <NC> GLOBal :CURRent :[AMPLitude] <NRf+> Set the output current for all power supplies :VOLTage :[AMPLitude] <NRf+> Set the output voltage for all power supplies :OUTPut:STATe <bool>...
Page 114
NATIONAL INSTRUMENTS :MODE OFF|FSTR|TRIG Set output trigger mode <NC> :RELay {1|2} [:STATe] <bool> Set control pins status REL{1|2}[?] :MODE? Returns the operation mode CV/CC/OFF MODE? [SOURce] :CURRent [:LEVel] [:IMMediate] [:AMPLitude] <NRf+> [:AMPLitude] <NRf+> Set the output current PC[?] :TRIGger <NRf+> Set the triggered output current <NC>...
Page 115
NATIONAL INSTRUMENTS :VOLTage <NRf+> Set output voltage points <NC> STATus :OPERation [:EVENt]? Returns the value of the Event register SEVE? :CONDition Returns the value of the Condition register STAT? :ENABle <NRf> Enables specific bits in the Event register SENA[?] :QUEStionable [:EVENt]? Returns the value of the Event register FEVE?
NATIONAL INSTRUMENTS CHAPTER 8: ADVANCED FUNCTIONS 8.1 Introduction This chapter describes the advanced functions of output programmable modes. The are three programmable modes: FIX, WAVE and LIST. The user can program the output transient state. The programmable modes are synchronized by input trigger (refer to section 8.5.1). According to the selected mode, the power supply activates the output trigger on J3-3 (refer to section 8.5.2).
NATIONAL INSTRUMENTS 8.3 LIST Mode Output change value in step determined by parameters in the LIST subsystem via input trigger. Refer to section 7.12.8 Fig.8-3: Simplified Models of LIST Mode 0.01sec Trig Initated INIT:IMM (Wait Trigger) Trigger Event 0.02s TRIG:DEL 0.02 (Step Started) DWELL 0.05s...
NATIONAL INSTRUMENTS 8.4 WAVE Mode Output change value in slope determined by parameters in the WAVE subsystem via input trigger. Refer to section 7.12.12. Fig.8-5: Simplified Models of Trigger in WAVE Mode 0.01sec Trig Initated INIT:IMM (Wait Trigger) Trigger Event 0.02s TRIG:DEL 0.02 (Step Started)
NATIONAL INSTRUMENTS 8.5 Trigger Subsystem Display Function Level Display Parameter Level Display Description Level Initialization INIT init INIT iniT (Refer to command INIT) TRIG TriG ready for trigger. refer to command Continue ConT INIT:CONT BUS (via Software Trigger Bs refer to command or front Panel), IN (input tr.
NATIONAL INSTRUMENTS 8.5.2 Output Trigger Rear panel connector J3-3 (Refer to section 4.3.2). There are three output trigger modes: Programming Mode NONE, FIX: • OFF – No Trigger out. • In TRIG mode, trigger is generated when output status changes. •...
NATIONAL INSTRUMENTS ready for trigger and not in an idle state, send ABORt and then set INIT:CONT 0. 8.6.3 Wave Execution via Front Panel 1. Load stored data MENU -> ”ROG” -> ”OAD” -> 2 2. Set Trigger IN source ( BUS for Command or Front Panel and EXT via Rear Panel J3.2 pin) MENU ->...
NATIONAL INSTRUMENTS CHAPTER 9: STATUS, FAULT AND SRQ REGISTERS 9.1 General This section describes various status errors (faults) and SRQ register structures. The registers can be read or set via the RS232/485/USB commands. Refer to Fig.9-1 for the Status and Fault Registers Diagram.
NATIONAL INSTRUMENTS 9.2 Power Supply Status Structure Status and Fault Register shows the status register structure of the power supply. The Standard Event, Status Byte, and Service Request Enable registers and the Output Queue perform standard functions as defined in the IEEE 488.2 Standard Digital Interface for Programmable Instrumentation. The Operation Status and Questionable Status registers implement status functions specific to the power supply.
NATIONAL INSTRUMENTS 9.3.2 Status Register The status register sets a bit when status changes (Refer to Table 9-2). The bit is cleared when the condition is removed. Bit Number Decimal Value Bit Symbol Description Set high if Constant Voltage Operation Set high if Constant Current Operation No fault Trigger wait...
NATIONAL INSTRUMENTS 9.6 Standard Event Status Group 9.6.1 Register Functions This group consists of an Event register and an Enable register that are programmed by COMMON commands. The Standard Event register latches events relating to interface communication status. It is a read-only register that is cleared when read. The Standard Event Enable register functions similarly to the Enable registers of the Operation and Questionable status groups.
NATIONAL INSTRUMENTS CHAPTER 10: MAINTENANCE 10.1 Introduction This chapter provides information about maintenance, calibration and troubleshooting. 10.2 Units Under Warranty Units requiring repair during the warranty period should be returned to a NI authorized service facility. Unauthorized repairs performed by other than the authorized service facilities may void the warranty.
NATIONAL INSTRUMENTS SYMPTOM CHECK ACTION REF. Check continuity, replace Is the AC power cord defective? if necessary. No output. All displays and indicators are blank. Is the AC input voltage Check input AC voltage. within range? Connect to appropriate voltage source. Output is present momentarily Does the AC source voltage sag Check input AC voltage.
Need help?
Do you have a question about the RMX-410 Series and is the answer not in the manual?
Questions and answers