Section 7 Troubleshooting; Troubleshooting The Controller; Troubleshooting The Analyzer; Led Status - Hach AMTAX sc User Manual

Ammonium analyzers
Hide thumbs Also See for AMTAX sc:
Table of Contents

Advertisement

Section 7

7.1 Troubleshooting the controller

7.2 Troubleshooting the analyzer

7.2.1 LED status

LED status
green LED
orange LED
LED flashing

7.2.2 Error messages

Error displayed
Instrument reaction
Warms up and goes
TEMP. < 0 °C/ 32 °F?
into the service state
Troubleshooting
If entries are only implemented with a delay or are not accepted for
a short time, the delay may be caused by a busy data network.
Refer to the troubleshooting section in the sc1000 User Manual.
If, in normal operation, problems occur that are apparently caused
by the controller, restart the system.
After a software update, a system expansion or after an interruption
in the power supply, it may be necessary to set the system
parameters again.
Note all the values that are changed or entered so all the necessary
data can be used to configure the parameters again.
1. Save all important data.
2. Isolate the supply of power and wait for 5 seconds.
3. Reapply power to the controller.
4. Check all relevant settings.
5. If the problem still occurs, contact Technical Support.
If the complete analysis instrument is not working, check whether
the humidity sensor has tripped. Fix the damage, dry the humidity
sensor and restart the system.
If the problems still occurs, contact Technical Support.
Table 9 LED status and definition
red LED
Cause
Instrument was below
4 °C (39 °F) on power
up
Definition
No errors of warnings
Error
Warning
No communication with controller
Solution
Reset error
Check whether
instrument is frozen
(Cleaning solution/
sample/reagent/
standards/electrode).
Reset error manually
If necessary use
TEST/MAINT>RESET
pre-warmed reagents.
ERROR
Thaw electrode, delete
error. Instrument will
then continue to warm
up and start
57