Insert Tubing And/Or Cables; Connect The Filter Probe To The Analyzer - Hach AMTAX sc User Manual

Ammonium analyzers
Hide thumbs Also See for AMTAX sc:
Table of Contents

Advertisement

Installation

3.5.3 Insert tubing and/or cables

1
Guide the tubing or cables through
the enclosure openings
(Figure
15).

3.5.4 Connect the Filter Probe to the analyzer

26
2
Push the plug from the top onto the
tubing or onto the cable.
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
DANGER
Disconnect power from the sc analyzer at the sc1000 before
removing the protective covers in the analyzer.
Important Note: Make sure that for the use of a filter probe, the
level of the water in which the filtration probe is submersed, is
below the level of the analyzer.
1. Open the enclosure door and secure with the door hook.
2. Open the analyzer panel.
3. Remove the two screws from the protective cover and remove
the cover (item 1,
4. Connect the earth ground (green/yellow) wire (item 9,
Figure 16, page
terminal (item 5,
5. Connect the power connector to the appropriate terminal
connection (items 4 and 11,
6. Remove the three screws securing the bottom panel cover.
(item 2,
Figure 14, page
7. Connect the data connector (item 10,
the main board (item 12,
8. Install all covers and panels.
9. Connect the white air tubing (item 8,
the Filter Probe sc to the air tubing connection on the analyzer
(Figure 16, page
10. For sample and drain line connections refer to
3
Pull down the plug with the tubing
or the cables. Seal any unused
entry with Plug #3.
Figure 14, page
24).
28) from the Filter Probe sc to the ground
Figure 16, page
28).
Figure 16, page
24). Remove the panel.
Figure 16, page
Figure 16, page
Figure 16, page
28).
28).
28) to
28).
28) from
A.4 , page
79.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents