Ammonium analyzers
Hide thumbs Also See for AMTAX sc:
Table of Contents

Advertisement

1
Overflow vessel
2
Fitting of sample tube
3
Blind plug
The validation is finished when the process shows the
service mode and the remaining time is 0 seconds.
For the adjusted number of validation measurements, the
values are listed and the calculated average of this value is
displayed.
Note: The event log records the validation values and average value from
the analyzer.
10. Press ENTER to proceed.
Note: Press START to confirm the inquiry to return to the
measurement process or to the service mode.
11. Select ENTER and modify the instrument to the original
analyzer configuration.
12. Start the measurement mode or hold the service mode.
Figure 21 Modification of AMTAX sc
4
Valve block
5
Sample tube
Maintenance
53