Operation
5.2 Sensor setup menu (continued)
CONFIGURE (continued)
DISCHARGE VAL 1
LOCATION 2
EDIT NAME
SET PARAMETER
SELECT UNITS
QUANTITY CH 2
DISCHARGE VAL 2
MEASURING
SET INTERVAL
START BY BUS:
START BY BUS:
NUMBER OF MEAS.: Number of measurements after a bus activation.
DISCHARGE:
AVERAGE:
CLEANING
SET INTERVAL
START
DISCHARGE
SET OUTMODE
CUVETTE TEMP.
TUBE HEATING
ON
OFF
REAG. WARNING
REAG. WARNING
WARNING
STAT. MODUL.WAR.
40 %, 30 %, 15 %
STATUS MODUL.ERR
14 %,10 %,8 %, OFF
38
Number of discarded values after switching from channel 1 to channel 2. Available with the
2-channel version.
Settings for location 2
Enter the name for the measuring location as required. Available with the 2-channel version.
Select output: ammonium or ammonium nitrogen. Available with the 2-channel version.
Select output: mg/L or ppm. Available with the 2-channel version.
Number of measurements in succession ( = measurements on channel 2 +
DISCHARGE VAL 2 channel 2). Available with the 2-channel version.
Number of discarded values after switching from channel 2 to channel 1. Available with the
2-channel version.
Enter measurement interval (Time interval between two measurements). ATTENTION with
filtration probe and 5 min operation: increased pump speed in the filtration probe, annual
filtration probe maintenance is necessary.
YES/NO; option whether the instrument is measuring continuously or measurements are
triggered by field-bus. "Fieldbus" must be activated under "TEST/MAINT". Instrument will
switch to 5 min interval when activated.
Number of discharged values that precede the measurements.
Number of measurements that are averaged. (Affects only the bus triggered measurements)
Number of hours between the cleanings
Start time for the cleaning (In case of more than one cleaning per day: Enter start time for the
first cleaning.)
Number of measured values that are discarded after a cleaning operation.
Value that is output during a cleaning operation and the following discarded values.
HOLD = last measured value, SET TRANSFER = value to be entered.
Temperature for cuvette and electrode
Recommended: Select 45 °C at air temperature: up to 35 °C, 50 °C at air temperature: up to
40 °C, 55 °C at air temperature: up to 45 °C, at 55 °C reduced accuracy and shortened
electrode service life may occur.
Probe tube heating switches on at the start of the selected month when using the Filter
Probe sc.
Probe tube heating switches off at the end of the selected month when using the Filter
Probe sc.
On/Off
When On is selected: determines the warning output if the reagent levels are low
Determines the level below which the reagent must drop to trigger the warning.
Warning is released: when a filter probe is installed and the status of filtration modules fall
below level
An error is generated: when a filter probe is installed and the status of filtration modules fall
below level. When switched to OFF, a deactivated sample detection will be switched to
"warning".