Hach AMTAX sc User Manual page 115

Ammonium analyzers
Hide thumbs Also See for AMTAX sc:
Table of Contents

Advertisement

Tag name
Register #
SET OUTMODE
40086
CAL.
DISCHARGE
40087
CLEAN.
SET OUTMODE
40088
CLE.
SET OUTMODE
40089
SER.
LOCATION2
40090
SET PARAMETER 2
40098
GAIN CORR. 2
40099
MEAS.UNITS 2
40101
HUMIDITY ANALY
40102
SOFTWARE
40103
PROBE
HUMIDITY PROBE
40105
PROCESS STATE
40107
LAST CALIBRAT.
40108
START BY BUS
40110
FIELDBUS
40111
NUMBER OF
40112
MEAS.
DISCHARGE BUS
40113
AVERAGE
40114
Table 13 Sensor Modbus registers (continued)
Data
Length
R/W
type
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
String
8
R/W
Unsigned
1
R/W
Integer
Float
2
R/W
Unsigned
1
R/W
Integer
Unsigned
1
R
Integer
Float
2
R
Unsigned
1
R
Integer
Unsigned
1
R
Integer
Time2
2
R
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Fieldbus Communications
Discrete
Min/Max
Description
range
range
Set output mode for calibration;
0/1
0=HOLD, 1= TRANSFER VALUE
0/10
Discharged values after a cleaning
Set output mode for cleaning;
0/1
0=HOLD, 1= TRANSFER VALUE
Set output mode for service mode;
0/1
0=HOLD, 1= TRANSFER VALUE
Location for the measuring channel
two for the sample
Select the parameter for channel 2;
19/42
19=NH4-N, 42=NH4
0.01/100
Gain correction for channel two
.00
Measurement units for channel 2;
0/2
0=mg/L, 2=ppm
0/100
Humidity analyzer in percent
0/3.4028
Software version of the filtration
2347E+
probe
38
0/100
Humidity filtration probe in percent
Process status is coded as enum
list; enum value 0=service mode,
enum value 1=citrocal in proc, cal1
in proc., cal2 in proc., measuring
0/1/2/3/4/
1..., interval, initialisation, serv.in
5/6/7/8/9/
proc., cleaning, warm up phase,
10/11/12/
measuring 2..., prepump reag.,
13/14/15/
prepump clean.,prepump standard,
16/17/18/
prepumping probe, flushing, start by
19/20
BUS, warming up, reserved,
prepump.sample; enum value
20=validation
Date of the last calibration
To set the analyzer to START BY
0/1
BUS mode (see Start by BUS)
To trigger a measurement serie in
0/2
START BY BUS mode (see Start by
BUS)
Number of measurements in a
1/100
START BY BUS measurement
series (see Start by BUS)
Discharge values at the beginning
3/10
of a START BY BUS series
Number of measurement values
that result in an average value in a
START BY BUS measurement
series.
115