5-9-2 Position NoDispense controls

The use of NoDispense controls or blanks were described above in
controls‖.
If any fluids selected in the Wizard were designated as ―NoDispense‖ fluids, this option will
1.
appear. Click on the ―Positioning: NoDispenses‖ in the left-hand menu. The window that
will appear is shown in
2.
Select which of the fluids you initially designated as NoDispense fluids you would like to
position in the well plate first, using the drop down menu titled ―NoDispense Fluid‖.
3.
Left-click and drag to designate wells for this NoDispense fluid. This Wizard window works
slightly differently than the main well plate image in that if you select any wells for this
NoDispense fluid by mistake, you have to right-click those wells to unselect them.
4.
Repeat steps 2 and 3 for each NoDispense fluid you wish to position. Click on Next when
finished, or select ―Positioning: Titrations‖ from the left-hand menu.
Figure 5-20
Positioning: NoDispenses pane
5-9-3 Position a single zone titration1
1.
Click the arrow on the drop-down list box to select one of the four positioning preferences
that the 96 points can be placed in.
concentrations in high to low order across the rows.
Figure
5-20.
Figure 5-21
HP D300 Digital Dispenser
―Position NoDispense
shows the layout by rows with the
47