Decontamination - Beckman Coulter SX4400 Instructions For Use Manual

Swinging bucket rotor
Table of Contents

Advertisement

2
Wash the buckets, covers, and O-rings with a mild detergent such as Solution 555, diluted 10 to
1 with water, and a soft brush.
3
Thoroughly rinse with water and air-dry.
4
Replace the O-ring in the groove.
Adapters
1
Disassemble modular adapters for washing by pulling them apart.
2
Use a mild detergent such as Solution 555, diluted 10 to 1 with water, and a soft brush to scrub
the adapters.
3
Rinse thoroughly with water.
4
Air-dry upside down.
• Reassemble modular adapters.

Decontamination

If aluminum rotor components become contaminated with radioactive material, decontaminate
them using a solution that will not damage the anodized surfaces. Beckman Coulter has tested a
number of solutions and found two that do not harm anodized aluminum: RadCon Surface Spray or
IsoClean Solution (for soaking),
While Beckman Coulter has tested these materials and found that they do not damage components,
no guarantee of decontamination is expressed or implied. Follow appropriate decontamination
procedures as directed by your laboratory safety officer.
* In U.S., contact Nuclear Associates (New York); in Eastern Europe and Commonwealth States, contact Victoreen GmbH
(Munich); in South Pacific, contact Gammasonics Pty. Ltd. (Australia); in Japan, contact Toyo Medic Co. Ltd. (Tokyo).
† In U.S., contact Biodex Medical Systems (Shirley, New York); internationally, contact the U.S. office to find the dealer
closest to you.
B01198AB
*
and Radiacwash.
SX4400 Swinging Bucket Rotor
Care and Maintenance
13