Sx4400 Swinging Bucket Rotor; Specifications - Beckman Coulter SX4400 Instructions For Use Manual

Swinging bucket rotor
Table of Contents

Advertisement

Specifications

r min
(71 mm)
r av
(101 mm)
r max
(172 mm)
1
1. Axis of Rotation
a. The critical speed range is the range of speeds over which the rotor shifts so as to rotate about its center of mass. Passing through the critical
speed range is characterized by some vibration.
b. Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius and speed (rw
gravity (g) according to the following formula: RCF = rw2/g — where r is the radius in millimeters, w is the angular velocity in radians per second
(2 p RPM /60), and g is the standard acceleration of gravity (9807 mm/s
B01198AB

SX4400 Swinging Bucket Rotor

Maximum speed
Allegra X-30R 4700 rpm
Allegra X-30 4200 rpm
Density rating at maximum speed 1.2 g/mL
a
Critical speed range
800 to 1200 rpm
b
Relative Centrifugal Field
at maximum speed at r
Allegra X-30R 4255 ´ g
Allegra X-30 3398 ´ g
Conditions requiring speed reductions see
Maximum allowable imbalance of opposing loads 10 grams
Number of buckets 4
Available labware
Table 1
Nominal tube capacity (largest tube) 400 mL
Nominal rotor capacity 1600 mL
Approximate acceleration time to maximum speed
(fully loaded) 36 sec
Approximate deceleration time from maximum speed
(fully loaded) 29 sec
Weight of fully loaded rotor 5.7 kg (12.6 lb)
Rotor and bucket material anodized aluminum
Bucket cover material polysulfone
2
). After substitution: RCF = 1.12r (RPM/1000)
(172 mm)
max
Run Speeds
2
) to the standard acceleration of
2
1

Hide quick links:

Advertisement

Table of Contents
loading